Adjustable buttons for ACL graft cortical fixation partially fail with cyclic loading and unloading
J Glasbrenner1, C Domnick1, M J Raschke1
1Department of Trauma, Hand and Reconstructive Surgery, Westphalian Wilhelms University, Waldeyer Strasse 1, 48149, Muenster, Germany.
Adjustable button devices for anterior cruciate ligament (ACL) reconstruction using the Chinese finger trap technique showed greater displacement under cyclic loading and unloading than other fixation methods. This suggests potential graft fixation failure, highlighting the need for careful rehabilitation or alternative fixation designs.
Area of Science:
- Orthopedic surgery
- Biomedical engineering
- Sports medicine
Background:
- Adjustable button (AB) devices are used for anterior cruciate ligament (ACL) soft-tissue graft fixation.
- The Chinese finger trap technique is a common mechanism in some AB devices.
- The cyclic loading performance, especially with complete unloading, of these devices is not well understood.
Purpose of the Study:
- To evaluate the cyclic loading resistance of AB devices utilizing the Chinese finger trap technique.
- To compare the displacement of Chinese finger trap AB devices against other fixation methods under cyclic loading with complete unloading.
- To test the hypothesis that Chinese finger trap AB devices exhibit greater displacement than continuous suture loop devices and other AB mechanisms.
Main Methods:
- Anterior cruciate ligament (ACL) reconstruction was performed in a porcine knee model.
- Three types of cortical fixation devices were tested: two Chinese finger trap AB devices, one locked suture loop AB device, and two continuous loop devices.
- Specimens underwent 2500 cycles of loading (up to 250 N) and complete unloading in a material-testing machine, with continuous elongation recorded.
Main Results:
- AB devices using the Chinese finger trap loop showed significantly greater displacement (mean 8.1 mm and 6.1 mm) compared to AB devices with a locked suture loop (mean 4.7 mm) and continuous loop devices (mean 4.1 mm and 4.4 mm).
- No significant difference in displacement was found between the locked suture loop AB devices and the continuous loop devices.
- The study demonstrated that Chinese finger trap AB devices are susceptible to construct lengthening under cyclic loading and unloading.
Conclusions:
- Adjustable button devices employing the Chinese finger trap technique exhibit significantly greater lengthening under cyclic loading and unloading compared to other fixation methods.
- Complete unloading of the ACL graft is a common occurrence during post-operative rehabilitation.
- The lengthening of Chinese finger trap AB devices may represent a potential failure mode for ACL graft fixation, suggesting consideration of locked suture loop designs or cautious rehabilitation protocols.
More Related Videos
Related Concept Videos
The Anchoring-and-Adjustment Heuristic
Hypothesis: Accept or Fail to Reject?
There are two ways to indicate that the null hypothesis is not rejected. 'Accept' the null...
Adjusting a Traverse
Cardiopulmonary Resuscitation II: ACLS Airway Management
Partial Fractions
Fixation and Sectioning
The simplest type of preparation is the wet mount, in which the specimen is placed in a drop of liquid on the slide. A liquid specimen can be directly deposited on the slide using a dropper. Solid specimens, such as skin scraping, can be placed on the slide before adding a drop of liquid to prepare the wet mount. Sometimes the liquid is simply water, but stains are often added...