Objective:
The Nucline X-Ring 4R is a four-headed gamma camera dedicated to neuroimaging. In this paper, we describe and validate a GATE (Geant4 Application for Tomographic Emission) model of the Nucline X-Ring 4R.
Materials And Methods:
Images produced during model simulations were compared with those acquired experimentally to confirm the model was an accurate representation of the scanner. The most commonly reported measurements used to validate a GATE model include energy resolution, spatial resolution and sensitivity. In addition to the commonly reported static imaging measures, single-photon emission computed tomography (SPECT) spatial resolution was investigated to confirm that the model produces similar SPECT images to the experimental output.
Results:
The experimental full-width at half-maximum was calculated to be 12.3 keV, which corresponds to an energy resolution of 8.8%. The simulated full-width at half-maximum was measured to be 12 keV, giving an energy resolution of 8.6%. The average spatial resolutions were found to be well matched (5.69 mm - simulated and 5.64 mm - experimental). However, the sensitivity was overestimated using the GATE model (47.8 and 54.3 cps/MBq) compared with the values obtained experimentally (42.7 and 44.3 cps/MBq). Finally, the simulated SPECT spatial resolution images were found to produce qualitatively comparable results.
Conclusion:
The model developed has been shown to produce similar results and images to those obtained experimentally. This model has the potential to simulate patient scans with the aim of improving patient care by optimizing scanner protocols.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies III: Computed Tomography
Emission Spectra
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Reliability and Validity