Factors influencing the accuracy of iliosacral screw insertion using 3D fluoroscopic navigation
Masaki Takao1, Hidetoshi Hamada2, Takashi Sakai2
1Department of Orthopaedic Medical Engineering, Osaka University Graduate School of Medicine, 2-2, Yamadaoka, Suita, Osaka, 565-0871, Japan. masaki-tko@umin.ac.jp.
Introduction:
The purpose of the present study was to determine which factors affect the positional accuracy of iliosacral screws inserted using 3D fluoroscopic navigation. Specifically, we asked: (1) does the screw insertion angle in the coronal and axial planes affect the positional accuracy of iliosacral screw insertion using 3D fluoroscopic navigation? (2) Is the positional accuracy of iliosacral screw insertion using 3D fluoroscopic navigation affected by the type of screw (transsacral versus standard iliosacral), site of screw insertion (S1 versus S2), patient position (supine versus prone), presence of a dysmorphic sacrum, or AO/OTA classification (type B versus C)?
Materials And Methods:
Twenty-seven patients with AO/OTA type B or C pelvic ring fracture were treated by percutaneous iliosacral screw fixation. A total of 55 screws were inserted into S1 or S2 using 3D fluoroscopic navigation combined with preoperative CT-based planning. The positional accuracy of screw placement was assessed by matching postoperative CT images with preoperative CT images. The distance between the central axis of the inserted screw and that of the planned screw placement was measured in the sagittal plane passing through the center of the vertebral body.
Results:
The mean deviation between the planned and the inserted screw position was 2.9 ± 1.7 mm (range 0-8.5 mm) at the vertebral body center. Multiple regression analysis showed that the screw insertion angle relative to the vertical line of the bone surface in the axial plane (β = 0.354, p = 0.013) and the use of a transsacral screw (β = 0.317, p = 0.017) were correlated with the positional accuracy of screw placement (adjusted R2 = 0.276, p = 0.002).
Conclusions:
A greater screw insertion angle relative to the vertical line on the bone surface and the use of transsacral screws increases the positional error of iliosacral screws inserted using 3D fluoroscopic navigation.
Level Of Evidence:
Level IV, therapeutic study.
Related Concept Videos
Bioavailability: Influencing Factors
Factors Influencing Microbial Growth: pH
Factors Influencing Heart Rate
Let us explore the significant factors affecting heart rate, including age, body temperature, posture, acute pain, chemical influences,...
Factors Influencing Attraction I: Proximity
Factors Influencing Attraction VI: Personality Traits
Factors Influencing Bioavailability: First-Pass Elimination