Performance of blow-by methods in delivering oxygen to pediatric patients during transport: A laboratory study
Clemens R M Barends1, Pembegul Yavuz1, Bouwe Molenbuur1
1Department of Anesthesiology, University of Groningen and University Medical Center Groningen, Groningen, The Netherland.
Background:
Providing supplemental oxygen with a blow-by method is used to provide additional oxygen to patients who will not tolerate an oxygen delivery device in direct contact with their face. Blow-by methods are often improvised from parts of standard equipment. The performance is very dependent on the distance to the face and the direction of the gas flow. Blow-by methods are used by anesthetists during transport but their performance in delivering supplemental oxygen has only been tested in static situations. The aim of this nonclinical study was to determine the performance of different blow-by methods in the delivery of additional oxygen to pediatric patients during transport.
Methods:
A manikin of a child with a facemask of appropriate size was transported along a 60 m corridor from the operating theater to the PACU. Oxygen delivery to the face of the manikin was measured during transport. Six blow-by methods were tested with oxygen flows of 3, 6, and 10 L/min and with the facemask at 0 cm from the face and at 5 cm from the face. The outcome parameter was: blow-by method reaching and maintaining an FiO2 >50% during transport from the pediatric operating theater to the PACU.
Results:
At 0 cm from the face, five out of six blow-by methods maintained a FiO2 >50% with all three flow rates. At 5 cm only two of the blow-by methods were able to maintain an FiO2 >50% and this only at flow rates of 10 L/min. All other blow-by methods provided lower FiO2 s; in three, the FiO2 decreased to values only marginally above 21%. The decrease in FiO2 typically started within 6-12 m from the start of the transport.
Conclusion:
It is concluded that the ability of blow-by methods to deliver a FiO2 >50% depends on the method used and distance from the face.
Related Concept Videos
Oxygen Transport in the Blood
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Pharmacokinetics in Pediatric Patients: Drug Excretion
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Pharmacokinetics in Pediatric Patients: Drug Distribution