The Symptoms and Clinical events associated with Automatic Reprogramming (SCARE) at replacement notification study
Sunil K Sinha1, Daniel Carlson1, Jonathan Chrispin1
1Division of Cardiology, Johns Hopkins University School of Medicine, Baltimore, MD, USA.
Background:
Pacemaker patients experience battery depletion that activates pacemaker's alert for replacement notification. Automatic reprogramming at replacement notification can result in loss of rate response and atrioventricular (AV) synchrony.
Objective:
To determine if relevant symptoms or clinical events may be associated with automatic reprogramming at replacement notification.
Methods:
Electronic medical record review was undertaken for 298 patients referred for pacemaker generator replacement. Primary endpoints were symptoms or clinical events during replacement notification period.
Results:
Following elimination of duplicate pacemaker replacements (n = 12), "near-replacement notification" or "recalled" (n = 15) and pacemakers at "end of life" (n = 5), 266 subjects were included. Three distinct reprogramming cohorts were identified; those with no change (control) in pacing mode (n = 46), those with loss of rate response (n = 154), and those with loss of AV synchrony ± rate response (n = 66). In total, 83 subjects (31.2%) had symptoms with significant differences seen between groups (control = 4.3%, loss of rate response = 26.0%, loss of AV synchrony ± rate response = 62.1%, P < 0.001). Overall, 28 subjects (10.5%) experienced clinical events with significant differences seen between groups (control = 0.0%, loss of rate response = 6.5%, loss of AV synchrony ± rate response = 27.3%, P < 0.001).
Conclusions:
Automatic reprogramming at replacement notification was associated with significant symptoms in 26% of those who lost rate response and in 62% of those who lost AV synchrony ± rate response. Additionally, 27% of the latter cohort required nonelective clinical care.
Related Concept Videos
Automatic Processing and Automatic Social Behavior
Aneurysm II: Clinical Manifestations and Diagnostic Studies
Diabetes: Symptoms, Diagnosis, and Complications
Negative and Cognitive Symptoms of Schizophrenia
Negative Symptoms
Negative symptoms of schizophrenia manifest as deficits in normal emotional and behavioral functioning, profoundly impacting daily life. Individuals with schizophrenia often display a flat affect, characterized by a near-total absence of emotional expression,...
Hypertension III: Clinical Manifestations and Diagnostic Studies
Venous Thrombosis II: Clinical Manifestations and Diagnostic Studies