Biomechanical Comparison of 3 Syndesmosis Repair Techniques With Suture Button Implants
Andrew S Parker1, David P Beason1, Jonathan S Slowik1
1American Sports Medicine Institute, Birmingham, Alabama, USA.
Background:
Suture button fixation of syndesmotic injury is growing in popularity, as it has been shown to provide adequate stability in a more cost-effective manner than screw fixation while allowing more physiologic distal tibiofibular joint motion. However, the optimal repair technique and implant orientation have yet to be determined.
Purpose/Hypothesis:
The purpose of this study was to biomechanically compare 3 suture button construct configurations/orientations for syndesmosis fixation: single, parallel, and divergent. The authors hypothesized that all 3 methods would provide adequate stabilization but that the divergent technique would be the most stable.
Study Design:
Controlled laboratory study.
Methods:
The fixation strengths of 3 stabilization techniques with suture button devices were compared with 10 cadaveric legs each (N = 30). Ankle motion under cyclic loading was measured in multiple planes: first in the intact state, then following simulated syndesmosis injury, and then following fixation with 1 of 3 randomly assigned constructs-1 suture button, 2 suture buttons in parallel, and 2 divergent suture buttons. Finally, axial loading with external rotation was applied to failure.
Results:
All syndesmotic fixation methods provided stability to the torn state. There was no statistically significant difference among the 3 fixation techniques in biomechanical stability. Failure most commonly occurred through fibular fracture at supraphysiologic loads.
Conclusion:
Suture button implant fixation for syndesmotic injury appears to provide stability to the torn syndesmosis, and the configuration of the fixation does not appear to affect the strength or security of the stabilization.
Clinical Relevance:
This study provides further insight into the biomechanics and optimal configuration of suture button fixation of the torn syndesmosis. Based on these results, the addition of a second suture button may not significantly contribute to immediate postoperative stability.
More Related Videos
Related Concept Videos
Mismatch Repair
Mismatch Repair
The Mutator Protein Family Plays a Key Role in DNA Mismatch Repair
The human genome has more than 3 billion base pairs of DNA per cell. Prior to cell division, that vast amount of genetic...
Overview of DNA Repair
Chemically...
Base Excision Repair
The first step of...
Nucleotide Excision Repair
Sutures of the Skull
Sutures are immobile joints between adjacent bones of the skull. The narrow gap between the bones is filled with dense, fibrous connective tissue that unites the bones. The long sutures located between the skull bones are not straight but instead follow irregular, tightly twisting paths. These twisting lines tightly...