Published on: October 10, 2020
Evaluation of the milling technique on the component-bone interface in the TATE elbow
Mark J Gosling1, Fernando Martinez-Taboada2, Gerard Te Lintelo1
1Wear Referrals, Bradbury, Stockton-on-Tees, United Kingdom.
Objective:
To determine whether the surgeon can influence the accuracy of milling during the TATE elbow arthroplasty by applying force to the milling arm during the milling procedure.
Study Design:
In vitro study on Sawbone specimens.
Sample Population:
Thirty identical solid foam models of the canine right humerus, elbow joint, and antebrachium (size large).
Methods:
TATE elbow arthroplasty was performed on 30 elbows equally divided into 3 types of forces applied to the milling arm: (1) no force, (2) a maximally converging force, and (3) a maximally diverging force using the center of rotation post as a reference point. The resulting component-bone interface and post fit were quantified with digital photography.
Results:
The component-bone interface gap differed between techniques for most angles. The application of convergent or divergent milling forces frequently increased the component-bone interface gap. Post fit was also influenced by the milling technique, application of a convergent or divergent milling force affecting the fit of multiple posts. Interface gaps tended to be greater on the lateral aspect, an area that is not visible to the surgeon intraoperatively.
Conclusion:
In this bone model study, application of convergent or divergent forces during milling frequently resulted in greater component-bone interface gaps and poorer post fit than when a neutral position was maintained.
Clinical Significance:
Surgeons may affect implant fit within the constraints of the current arthroplasty system if they do not maintain a neutral position during milling. Greater component-bone interface gaps and poorer post fit may affect component osseointegration and lead to aseptic implant loosening.
More Related Videos
11:03Nanoscale Characterization of Liquid-Solid Interfaces by Coupling Cryo-Focused Ion Beam Milling with Scanning Electron Microscopy and Spectroscopy
Published on: July 14, 2022
06:59Author Spotlight: An Economic and Efficient Method for Quantitative Evaluation of Bone Microarchitecture in a Murine Osteoporosis Model
Published on: September 8, 2023
Related Concept Videos
Protein-protein Interfaces
Protein-Protein Interfaces
Components of Stress
Interestingly, the hidden cube faces also experience these stresses, equal and...
Components of Language
Vector Algebra: Method of Components
In many applications, the magnitudes and directions of...
Curvilinear Motion: Rectangular Components
As the car advances, its position evolves over time. Quantifying the car's velocity involves computing the...