Influence of Experimental Conditions on the Ratio of 25-Hydroxyvitamin D3 Conformers for Validating a Liquid
Nicholas R Oranzi1, Nicolas C Polfer1, Jiajun Lei1
1Department of Chemistry , University of Florida , Gainesville , Florida 32611 , United States.
Abstract:
The utility of adding ion mobility (IM) to liquid chromatography/mass spectrometry (LC/MS) for quantitation of 25-hydroxyvitamin D3 (25OHD3) is evaluated. Sodiated 25OHD3 ions adopt both closed and open conformations, whereas the stereoisomer 3-epi-25-hydroxyvitamin D (epi25OHD), when sodiated, adopts only a closed gas-phase conformation. The unique open conformation for sodiated 25OHD3 permits unambiguous quantitation. Nonetheless, the ratio of open versus closed gas-phase conformations for sodiated 25OHD3 can vary with instrumental conditions; conversion from the open to the closed conformer is attributed to radio frequency (rf) heating within the ion accumulation trap. Ion heating becomes significant when space charge in the trap forces ions to larger radii where the rf field is higher. To avoid biasing quantitation, an isotopically labeled internal standard must be used to account for changes in conformer ratio. Thirty-three serum extract samples were evaluated by LC/IM-MS and were found to not be biased by changes in ion conformer ratios, permitting reliable quantitation of 25OHD3 without interference from the epimer.
More Related Videos
08:56Quantitative Analysis of the Cellular Lipidome of Saccharomyces Cerevisiae Using Liquid Chromatography Coupled with Tandem Mass Spectrometry
Published on: March 8, 2020
09:09Semi-Quantitative Analysis of Peptidoglycan by Liquid Chromatography Mass Spectrometry and Bioinformatics
Published on: October 13, 2020
Related Concept Videos
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Conformity
Ion-Exchange Chromatography
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Instrumentation
Mass Spectrometry: Overview