Measuring the Carotid to Femoral Pulse Wave Velocity Cf-PWV to Evaluate Arterial Stiffness
Published on: May 3, 2018
Automatic or manual arterial path for the ankle-brachial differences pulse wave velocity
Juan Francisco Sánchez Muñoz-Torrero1,2, Julián Fernando Calderón-García2,3, Jorge Manuel De Nicolás-Jiménez2,4
1Department of Internal Medicine, Hospital San Pedro de Alcántara, Extremadura Health Service, Cáceres, Spain.
Automated arterial path length measurements differ from manual methods, impacting pulse wave velocity calculations. Height-based formulas may be recommended for ankle-brachial difference in pulse wave velocity (abD-PWV) assessment.
Area of Science:
- Cardiovascular physiology
- Medical device technology
- Biomedical engineering
Background:
- Automated methods for measuring arterial path length are common for pulse wave velocity (PWV) assessment.
- Accurate arterial path length is crucial for reliable PWV measurements in peripheral arteries.
Purpose of the Study:
- To compare manual arterial path length measurements with height-based formulas.
- To evaluate the comparability of ankle-brachial difference in pulse wave velocity (abD-PWV) using automated versus manual methods.
Main Methods:
- Manual arterial path length measurement using a tape measure from suprasternal notch to VOPITB cuff midpoint in 245 patients.
- Comparison of manual measurements with height-based formulas.
- Assessment of abD-PWV using both manual and automated methods.
Main Results:
- Manual measurements yielded shorter arterial path lengths than automated calculations (13% arms, 21% legs).
- Automated abD-PWV was significantly higher (462 cm/s) than manual (346 cm/s).
- Bland-Altman analysis indicated percentage errors of 15.2% (heart-brachial), 7.5% (heart-ankle length), and 17.3% (abD-PWV), suggesting comparability despite differences.
Conclusions:
- Significant discrepancies exist between manual and automated arterial length measurements.
- These differences influence abD-PWV calculations.
- Height-based formulas may be the preferred method for abD-PWV measurement due to their advantages.
Related Concept Videos
Velocity and Acceleration of a Wave
The velocity of the particles can be obtained by taking the partial derivative of the position equation with respect to time....
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Mean free path and Mean free time
Ankle Joint
Automatic Processing and Automatic Social Behavior