Commissioning of a respiratory gating system involving a pressure sensor in carbon-ion scanning radiotherapy
Hideyuki Mizuno1, Osami Saito1, Minoru Tajiri1
1National institute of Radiological Sciences, QST, Chiba, Japan.
Abstract:
This study reports the commissioning methodology and results of a respiratory gating system [AZ - 733 V/733 VI (Anzai Medical Co., Japan)] using a pressure sensor in carbon-ion scanning radiotherapy. Commissioning includes choosing a location and method for pressure sensor installation, delay time measurement of the system, and the final flow test. Additionally, we proposed a methodology for the determination of a threshold level of generating an on/off gate for the beam to the respiratory waveform, which is important for clinical application. Regarding the location and method for installation of the pressure sensor, the actual person's abdomen, back of the body position, and supine/prone positioning were checked. By comparing the motion between the pressure sensor output and the reference LED sensor motion, the chest rear surface was shown to be unsuitable for the sensor installation, due to noise in the signal caused by the cardiac beat. Regarding delay time measurement of the system, measurements were performed for the following four steps: (a). Actual motion to wave signal generation; (b). Wave signal to gate signal generation; (c). Gate signal to beam on/off signal generation; (d). Beam on/off signal to the beam irradiation. The total delay time measured was 46 ms (beam on)/33 ms (beam off); these were within the prescribed tolerance time (<100 ms). Regarding the final flow test, an end-to-end test was performed with a patient verification system using an actual carbon-ion beam; the respiratory gating irradiation was successfully performed, in accordance with the intended timing. Finally, regarding the method for determining the threshold level of the gate generation of the respiration waveform, the target motion obtained from 4D-CT was assumed to be correlated with the waveform obtained from the pressure sensor; it was used to determine the threshold value in amplitude direction.
More Related Videos
08:10Precision Milling of Carbon Nanotube Forests Using Low Pressure Scanning Electron Microscopy
Published on: February 5, 2017
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
Related Concept Videos
Non-gated Ion Channels
Compared to the gated ion channels, the non-gated channels, also known as leakage or passive channels, have no gating mechanism....
Mechanically-gated Ion Channels
Ligand-gated Ion Channels
Three Subfamilies of Ligand-gated Ion Channels
Ligand-gated ion channels fall into three subfamilies. The 'Cys-loop' includes the nicotinic acetylcholine receptors, γ-aminobutyric acid (GABA), glycine, and 5-hydroxytryptamine receptors. The second one is the 'Pore-loop' channels that...
Voltage-gated Ion Channels
Generally, all voltage-gated ion channels have a 'voltage-sensing domain' that spans the lipid bilayer. The charged residues in the sensor move in response to the membrane potential changes that open the channel allowing ions movement. There are several types of...
Ligand-Gated Ion Channel Receptor: Gating Mechanism
G-Protein Gated Ion Channels
Sensory...