Concentric or monopolar electrode for jitter determination in orbicularis oculi
Miguel Oliveira Santos1, Michael Swash2, Mamede de Carvalho1
1Institute of Physiology Unit, Instituto de Medicina Molecular, Faculty of Medicine, University of Lisbon, Portugal; Department of Neurosciences and Mental Health, Hospital de Santa Maria, Centro Hospitalar Lisboa Norte, Portugal.
Monopolar (MN) and concentric (CN) electrodes are equally effective for estimating volitional contraction jitter in the orbicularis oculi (OO) muscle. Filter settings did not impact diagnostic accuracy for ocular myasthenia gravis (MG).
Area of Science:
- Neurology
- Neurophysiology
- Ophthalmology
Background:
- Ocular myasthenia gravis (MG) diagnosis relies on assessing neuromuscular function.
- Estimating volitional contraction jitter in the orbicularis oculi (OO) muscle is crucial for diagnosing MG.
- The choice of electrode type and filter settings may influence jitter measurements.
Purpose of the Study:
- To compare the efficacy of monopolar (MN) and concentric (CN) electrodes for jitter estimation in the OO muscle.
- To evaluate the impact of high-pass filter settings (1 kHz vs. 3 kHz) on jitter analysis.
- To determine diagnostic sensitivity and specificity for ocular MG using different electrode and filter configurations.
Main Methods:
- A randomized protocol was used to study 100 patients with suspected ocular MG.
- Neuromuscular jitter was assessed in the OO muscle using either MN (50 patients) or CN (50 patients) electrodes.
- Each patient had jitter measured on both sides, with one side using a 1 kHz high-pass filter and the other using a 3 kHz filter.
Main Results:
- In control subjects (n=52), jitter distribution was normal and unaffected by electrode type or filter settings.
- Mean jitter values and the proportion of abnormal jitter pairs were similar between MN and CN electrodes across both filter settings.
- Diagnostic sensitivity for ocular MG was 73% for mean jitter and 94% for abnormal pairs, with 100% specificity.
Conclusions:
- Both MN and CN electrodes are suitable and equivalent for assessing volitional jitter in the OO muscle.
- High-pass filter selection (1 kHz or 3 kHz) does not significantly alter the diagnostic utility of jitter measurements in ocular MG.
- These findings support the interchangeable use of MN and CN electrodes in clinical practice for OO jitter analysis.
Related Concept Videos
Concentration Cells
Consider the following voltaic cell:
Standard Electrode Potentials
Determination of Multiple Dosing Parameters: Steady-State, Minimum and Maximum Concentrations
Calculating Equilibrium Concentrations
A more...
Concentration and Rate Law
For example, in a generic reaction aA + bB ⟶ products, where a and b are stoichiometric coefficients, the rate law can be written as:
Solution Concentration and Dilution