Real-time scatter estimation for medical CT using the deep scatter estimation: Method and robustness analysis with
Joscha Maier1,2, Elias Eulig1,2, Tim Vöth1,2
1German Cancer Research Center (DKFZ), Im Neuenheimer Feld 280, 69120, Heidelberg, Germany.
Purpose:
X-ray scattering leads to CT images with a reduced contrast, inaccurate CT values as well as streak and cupping artifacts. Therefore, scatter correction is crucial to maintain the diagnostic value of CT and CBCT examinations. However, existing approaches are not able to combine both high accuracy and high computational performance. Therefore, we propose the deep scatter estimation (DSE): a deep convolutional neural network to derive highly accurate scatter estimates in real time.
Methods:
Gold standard scatter estimation approaches rely on dedicated Monte Carlo (MC) photon transport codes. However, being computationally expensive, MC methods cannot be used routinely. To enable real-time scatter correction with similar accuracy, DSE uses a deep convolutional neural network that is trained to predict MC scatter estimates based on the acquired projection data. Here, the potential of DSE is demonstrated using simulations of CBCT head, thorax, and abdomen scans as well as measurements at an experimental table-top CBCT. Two conventional computationally efficient scatter estimation approaches were implemented as reference: a kernel-based scatter estimation (KSE) and the hybrid scatter estimation (HSE).
Results:
The simulation study demonstrates that DSE generalizes well to varying tube voltages, varying noise levels as well as varying anatomical regions as long as they are appropriately represented within the training data. In any case the deviation of the scatter estimates from the ground truth MC scatter distribution is less than 1.8% while it is between 6.2% and 293.3% for HSE and between 11.2% and 20.5% for KSE. To evaluate the performance for real data, measurements of an anthropomorphic head phantom were performed. Errors were quantified by a comparison to a slit scan reconstruction. Here, the deviation is 278 HU (no correction), 123 HU (KSE), 65 HU (HSE), and 6 HU (DSE), respectively.
Conclusions:
The DSE clearly outperforms conventional scatter estimation approaches in terms of accuracy. DSE is nearly as accurate as Monte Carlo simulations but is superior in terms of speed (≈10 ms/projection) by orders of magnitude.
More Related Videos
Related Concept Videos
What are Estimates?
The estimate for the mean of a sample is denoted by ͞x, whereas the mean of the population is designated as μ. Further, parameters such...
Scatter Plot
Estimation of k and VD of Aminoglycosides
Truncation in Survival Analysis
Left truncation occurs when individuals who experienced the event of interest before a certain time are not included in the study. This is often due to a "delayed entry" into the study where only those who survive until a certain entry point are...
Estimation of the Physical Quantities
Estimating Population Standard Deviation