Evaluation of accuracy of automatic out-of-plane respiratory gating for DCEUS-based quantification using principal
Diya Wang1, Zhe Su2, Qiang Su3
1The Key Laboratory of Biomedical Information Engineering of Ministry of Education, Department of Biomedical Engineering, School of Life Science and Technology, Xi' an Jiaotong University, Xi'an, PR China; Laboratory of Biorheology and Medical Ultrasonics, University of Montreal Hospital Research Center, Montreal, QC, Canada.
Abstract:
The accuracy of abdominal multi-parametric quantification based on dynamic contrast-enhanced ultrasound (DCEUS) is limited by out-of-plane severe distortion induced by respiratory motion. This study developed a fully automatic respiratory gating scheme by using principal component analysis to remove distortions and disturbances in free-breathing DCEUS-based quantification. Taking the known in-vitro perfusions as ground truths, we further evaluated the respiratory gating accuracy from multiple perspectives in a controllable rotary distortion flow model with out-of-plane severe distortion induced by respiratory motion. Compared with those without respiratory gating, the signal-to-clutter ratio and correlation coefficients of perfusion parameters with respiratory gating improved by 3.99 ± 1.71 dB (p < 0.01) and 0.39 ± 0.17 (p < 0.01), respectively. The corresponding mean square error decreased by 1893.9 ± 763.16 (p < 0.001). Results with respiratory gating were significantly consistent with the ground truths without respiratory motion disturbances. Quantitative results demonstrated the effective removal of out-of-plane serious distortion and other disturbances induced by the respiratory kinetics on DCEUS-based quantification. The accuracy and robustness of DCEUS-based quantification were significantly improved by the proposed respiratory gating strategy, which not only benefits physicians in identifying transient hemodynamic characteristics in tumor angiogenesis, but also provides accurate diagnoses for abdominal tumors.
More Related Videos
07:51Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
08:08Evaluating Targeting Accuracy in the Focal Plane for an Ultrasound-guided High-intensity Focused Ultrasound Phased-array System
Published on: March 6, 2019
Related Concept Videos
Principal Stresses in a Beam
Analyzing principal stresses is crucial, especially in...
Improving Translational Accuracy
Automatic Processing and Automatic Social Behavior
Absolute Motion Analysis- General Plane Motion
As the drone's propellers rotate, an upward force is generated that counteracts the force of gravity, enabling the drone to lift off from the ground. This initial movement of the drone is along a straight path, representing a form of translational motion. In this phase, every point on the...
Principal Moments of Area
The principal moment of inertia axes are the...
Principal Stresses