Interatrial Shunting for Heart Failure: Early and Late Results From the First-in-Human Experience With the V-Wave
Josep Rodés-Cabau1, Mathieu Bernier1, Ignacio J Amat-Santos2
1Quebec Heart and Lung Institute, Quebec City, Quebec, Canada.
Insights
Interatrial shunting using the V-Wave device is a safe and feasible treatment for high-risk heart failure (HF). While shunt patency decreased over time, patients with open shunts showed improved clinical status and reduced hospitalizations.
Area of Science:
- Cardiology
- Medical Devices
- Heart Failure Research
Background:
- Heart failure (HF) remains a significant cause of morbidity and mortality, particularly in high-risk patients.
- Current therapies for advanced HF have limitations, necessitating novel treatment approaches.
Purpose of the Study:
- To assess the feasibility, safety, and exploratory efficacy of interatrial shunting with the V-Wave system in patients with high-risk HF.
- To evaluate the impact of the V-Wave shunt on clinical status, functional capacity, and hemodynamics.
Main Methods:
- A single-arm, open-label, first-in-human study involving 38 patients with New York Heart Association (NYHA) functional class III or IV HF.
- The V-Wave shunt was implanted via transseptal catheterization, with follow-up evaluations at 3, 12 months, and annually.
Main Results:
- Successful shunt implantation in all patients with no periprocedural mortality; major complications occurred in 2.6% of patients.
- Significant improvements observed in NYHA class, quality of life, and 6-minute walk distance at 3 and 12 months.
- While shunt patency declined by 12 months, widely patent shunts correlated with lower long-term mortality, need for advanced HF therapies, and HF hospitalizations.
Conclusions:
- Interatrial shunting with the V-Wave system is feasible and safe for high-risk HF patients.
- Early clinical and functional improvements were noted, though shunt patency durability requires further investigation.
- Device modifications to enhance shunt patency durability are recommended prior to randomized trials.
Objectives:
This was a first-in-human study to assess the feasibility, safety, and exploratory efficacy of interatrial shunting for treating high-risk heart failure (HF) in patients with reduced and preserved ejection fraction.
Methods:
A single-arm open-label study of patients with New York Heart Association functional class III or IV HF on optimal therapy was performed at 6 centers. The V-Wave shunt, an hourglass-shaped implant containing a 1-way bioprosthetic valve, was implanted by transseptal catheterization. Clinical, functional, echocardiographic, and hemodynamic evaluations were performed at baseline, 3 and 12 months, and annually (clinical follow-up) thereafter (median follow-up 28 months; interquartile range: 21 to 31 months).
Results:
A total of 38 patients were enrolled (30 with HF with reduced ejection fraction and 8 with HF with preserved ejection fraction; mean age 66 ± 9 years; 97% and 3% in New York Heart Association functional classes III and IV, respectively), and the shunt device was successfully implanted in all cases without periprocedural mortality. The rate of major device- or procedure-related complications during the first 12 months was 2.6% (periprocedural cardiac tamponade in 1 patient). At 3- and 12-month follow-up, there were improvements in New York Heart Association functional class (classes I and II in 78% and 60% of patients, respectively), quality of life (improvements ≥5 points in 74% and 73% of patients, respectively), and 6-min walk distance (mean increases of 41 ± 63 m and 28 ± 83 m, respectively) (p < 0.02 for all, data available for 36 patients), without changes in objective measures of left- or right-sided function. All shunts were patent at 3 months, but 5 of 36 (14%) had occluded, and another 13 of 36 (36%) were stenotic at the valve by 12 months. Patients with widely patent shunts had lower long-term rates of death, left ventricular assist device placement or heart transplantation (p = 0.001), and HF hospitalization (p = 0.008), along with a reduction of pulmonary capillary wedge pressure (from 23.3 ± 5.4 mm Hg at baseline to 18.0 ± 4.0 mm Hg at 12 months; p = 0.011).
Conclusions:
Interatrial shunting with the V-Wave system was feasible and safe in patients with HF with reduced and preserved ejection fraction. Improvements in clinical and functional status were observed early and at 12 months despite attenuation of shunt patency in one-half of the patients. Patients with preserved shunt patency tended to maintain clinical benefit during longer term follow-up. Device modification that improves the durability of patency is likely worthwhile before confirmation of these findings in a randomized trial.
More Related Videos
07:49Author Spotlight: Investigating HR-Dependent Cardiac Function in Mouse Models Through a Novel Atrial-Pacing Approach
Published on: July 21, 2023
09:00The Rabbit Blood-shunt Model for the Study of Acute and Late Sequelae of Subarachnoid Hemorrhage: Technical Aspects
Published on: October 2, 2014
Related Concept Videos
Heart Failure II: Pathophysiology
Pathophysiology of Heart Failure
Heart Failure I: Introduction
Heart Failure VI: Adjunct Therapies
Heart Failure Drugs: Diuretics
Heart Failure V: Medical Management