Delayed presentation of pulmonary artery perforation by an Amulet left atrial appendage closure device
Edward Wang1, Wah Wah Lin1, Xiao Fang Xu2
1Department of Cardiothoracic Surgery and Transplantation, Fiona Stanley Hospital, Murdoch, Western Australia, Australia.
Abstract:
Left atrial appendage occlusion (LAAO) devices offer stroke prevention in atrial fibrillation for patients intolerant of anticoagulation. Device placement leading to bleeding and cardiac tamponade have been reported periprocedurally but delayed presentations have not been reported in the literature. We present the case of an Amulet LAAO device causing erosion and bleeding from the main pulmonary artery that presented with cardiac tamponade 6 months after device placement. The pulmonary artery defect was repaired primarily and buttressed with a pericardial patch with good result.