Intra- and inter-operator reproducibility of automated cloud-based carotid lumen diameter ultrasound measurement
Luca Saba1, Sumit K Banchhor2, Tadashi Araki3
1Department of Radiology, University of Cagliari, Italy.
Background:
Common carotid artery lumen diameter (LD) ultrasound measurement systems are either manual or semi-automated and lack reproducibility and variability studies. This pilot study presents an automated and cloud-based LD measurements software system (AtheroCloud) and evaluates its: (i) intra/inter-operator reproducibility and (ii) intra/inter-observer variability.
Methods:
100 patients (83M, mean age: 68±11years), IRB approved, consisted of L/R CCA artery (200 ultrasound images), acquired using a 7.5-MHz linear transducer. The intra/inter-operator reproducibility was verified using three operator's readings. Near-wall and far carotid wall borders were manually traced by two observers for intra/inter-observer variability analysis.
Results:
The mean coefficient of correlation (CC) for intra- and inter-operator reproducibility between all the three automated reading pairs were: 0.99 (P<0.0001) and 0.97 (P<0.0001), respectively. The mean CC for intra- and inter-observer variability between both the manual reading pairs were 0.98 (P<0.0001) and 0.98 (P<0.0001), respectively. The Figure-of-Merit between the mean of the three automated readings against the four manuals were 98.32%, 99.50%, 98.94% and 98.49%, respectively.
Conclusions:
The AtheroCloud LD measurement system showed high intra/inter-operator reproducibility hence can be adapted for vascular screening mode or pharmaceutical clinical trial mode.
Related Concept Videos
Operational Amplifiers
Real Number Operations
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Design Example: Calculating Safe Diameter for Wind-Exposed Disc
Vector Operations
A vector multiplied by a scalar value is called scalar multiplication. The result obtained is a new vector with a different magnitude. If the scalar is positive, the direction of the vector remains the same, but if it is negative, the direction of the vector is reversed. For example, the product of the mass and velocity yields the momentum.
Operant Conditioning
Reinforcement in operant conditioning can be positive or negative, both of which serve to increase the likelihood of a behavior. Positive...