Cataract quantification using swept-source optical coherence tomography
João E G Brás1, Wolfgang Sickenberger1, Nino Hirnschall1
1From the SciTec Department (Brás, Sickenberger), JenVis Research GbR, Ernst-Abbe-Hochschule Jena, Jena, Germany Vienna Institute for Research in Ocular Surgery (Hirnschall, Findl), Ophthalmology Department, Hanusch-Hospital, Vienna, Austria.
Purpose:
To develop and evaluate a cataract quantification method using a swept-source optical coherence tomography (SS-OCT) device (IOLMaster 700).
Setting:
Hanusch Hospital, Vienna, Austria.
Design:
Prospective multicenter case series.
Methods:
This study included patients with cataract in at least 1 eye. Two independent examiners performed Lens Opacities Classification System II (LOCS II) grading at the slitlamp independently. Corrected distance visual acuity, contrast sensitivity, and SS-OCT measurements were also performed. In addition, the phacoemulsification energy and time were recorded. To develop an objective SS-OCT-based cataract quantification system, all SS-OCT scans were segmented and the local pixel intensity unit of the lens was analyzed using Matlab's grayscale counting. The pixel intensity unit of the lens nucleus (OCT-based cataract quantification system score) was equated to the clinician's subjective nuclear opalescence grading.
Results:
The study evaluated 186 eyes (113 patients). The correlation between the independent examiners' LOCS grading was good (0.91) (P < .01). The correlation between the LOCS grading and OCT-based cataract quantification system score was 0.86 for examiner 1 and 0.76 for examiner 2 (both P < .01). In 24 patients, phacoemulsification time, power, and energy; visual acuity; and contrast sensitivity were available and included in the study. The OCT-based cataract quantification system scores correlated significantly with phacoemulsification time (0.71) and energy (0.64) (both P < .01).
Conclusions:
Cataract density could be evaluated using an SS-OCT device, and the OCT-based cataract quantification system values correlated with the conventional LOCS II classification. Swept-source OCT measurements allowed quantification and documentation cataract density.
Related Concept Videos
Sinusoidal Sources
In homes, the power supplies use sinusoidal sources to provide electricity. These sources generate a voltage that varies sinusoidally...
AC Sources
Sources of Law
Constitutional law is foundational, deriving from federal and state constitutions, and...
Source Transformation
It is essential to note that when...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Multiple Voltage Sources
In series, the positive terminal of one battery is connected to the negative terminal of another battery. Hence, the voltage of each battery is added to give the net voltage, which is increased because each battery boosts the electrons that enter it. The same current flows through each battery because they are connected in series.
Batteries are...