Speed-adaptive control of functional electrical stimulation for dropfoot correction
Guangtao Chen1, Le Ma1, Rong Song2
1Key Laboratory of Sensing Technology and Biomedical Instrument of Guangdong Province, School of Engineering, Sun Yat-sen University, Guangzhou, Guang Dong, China.
Background:
Functional electrical stimulation is an important therapy technique for dropfoot correction. In order to achieve natural control, the parameter setting of FES should be associated with the activation of the tibialis anterior.
Methods:
This study recruited nine healthy subjects and investigated the relations of walking speed with the onset timing and duration of tibialis anterior activation. Linear models were built for the walking speed with respect to these two parameters. Based on these models, the speed-adaptive onset timing and duration were applied in FES-assisted walking for nine healthy subjects and ten subjects with dropfoot. The kinematic performance of FES-assisted walking triggered by speed-adaptive stimulation were compared with those triggered by the heel-off event, and no-stimulation walking at different walking speeds.
Results:
Higher ankle dorsiflexion angle was observed in heel-off stimulation and speed-adaptive stimulation conditions than that in no-stimulation walking condition at all the speeds. For subjects with stroke, the ankle plantarflexion angle in speed-adaptive stimulation condition was similar to that in no-stimulation walking condition, and it was significant larger than that in heel-off stimulation condition at all speeds.
Conclusions:
The improvement in ankle dorsiflexion without worsening ankle plantarflexion in speed-adaptive stimulation condition could be attributed to the appropriate stimulation timing and duration. These results provide evidence that the proposed stimulation system with speed-related parameters is more physiologically appropriate in dropfoot correction, and it may have great potential value in future clinical applications.
Trial Registration:
Medical Ethics Committee of Guangdong Work Injury Rehabilitation Center, AF/SC-07/2016.22 . Registered 26 May 2016.
Related Concept Videos
Distance Corrections
Transfer Function in Control Systems
To derive the transfer function, consider a general nth-order linear time-invariant...
Power Factor Correction
Combinatorial Gene Control
The expression of more than 30,000 genes is controlled by approximately 2000-3000 transcription factors. This is possible because a single transcription factor can recognize more than one regulatory sequence. The specificity in gene...
Speed of Sound in Gases
Speed of a Transverse Wave
One of the key properties of any wave is the wave speed. Light...