Related Experiment Video
Updated: Feb 2, 2026

Measurement of the Hepatic Venous Pressure Gradient and Transjugular Liver Biopsy
Published on: June 18, 2020
EUS-guided portal pressure gradient measurement with a simple novel device: a human pilot study
Jason B Samarasena1, Jason Y Huang1, Takeshi Tsujino1
1H. H. Chao Comprehensive Digestive Disease Center, Division of Gastroenterology and Hepatology, University of California, Irvine, Orange, California, USA.
Background And Aims:
Portal hypertension is a serious adverse event of liver cirrhosis. Recently, we developed a simple novel technique for EUS-guided portal pressure gradient (PPG) measurement (PPGM). Our animal studies showed excellent correlation between EUS-PPGM and interventional radiology-acquired PPGM. In this video we demonstrate the results of the first human pilot study of EUS-PPGM in patients with liver disease.
Methods:
EUS-PPGM was performed by experienced endosonographers using a linear echoendoscope, a 25-gauge FNA needle, and a novel compact manometer. The portal vein and hepatic vein (or inferior vena cava) were targeted by use of a transgastric or transduodenal approach. Feasibility was defined as successful PPGM in each patient. Safety was based on adverse events captured in a postprocedural interview.
Results:
Twenty-eight patients underwent EUS-PPGM with 100% technical success and no adverse events. PPG ranged from 1.5 to 19 mm Hg and had excellent correlation with clinical parameters of portal hypertension, including the presence of varices (P = .0002), PH gastropathy (P = .007), and thrombocytopenia (P = .036).
Conclusion:
This novel technique of EUS-PPGM using a 25-gauge needle and compact manometer is feasible and appears safe. Given the availability of EUS and the simplicity of the manometry setup, EUS-guided PPG may represent a promising breakthrough for procuring indispensable information in the management of patients with liver disease.
Related Concept Videos
Definition and Measurement of Pressure: Atmospheric Pressure, Barometer, and Manometer
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
Measurement of Blood Pressure
What is an Electrochemical Gradient?
The chemical gradient relies on differences in the abundance of a substance on the outside versus the inside of a cell and flows from areas of high to low ion concentration. In contrast, the electrical gradient revolves around an...
Pilot and Numeric Relaying
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...