Shear stress and blood trauma under constant and pulse-modulated speed CF-VAD operations: CFD analysis of the HVAD
Related Experiment Video
Updated: Feb 2, 2026

Introducing Shear Stress in the Study of Bacterial Adhesion
Published on: September 2, 2011
Zengsheng Chen1, Sofen K Jena2, Guruprasad A Giridharan2,3
1Artificial Organs Laboratory, Department of Surgery, University of Maryland School of Medicine, 10 South Pine Street, MSTF 434A, Baltimore, MD, 21201, USA.
Abstract:
Modulation of pump speed has been proposed and implemented clinically to improve vascular pulsatility in continuous flow ventricular assist device patient. The flow dynamics of the HVAD with a promising asynchronous pump speed modulation and its potential risk for device-induced blood trauma was investigated numerically. The boundary conditions at the pump inlet and outlet were defined using the pressure waveforms adapted from the experimentally recorded ventricular and arterial pressure waveforms in a large animal ischemic heart failure (IHF) model supported by the HVAD operated at constant and modulated pump speeds. Shear stress fields and hemolysis indices were derived from the simulated flow fields. The overall features of the computationally generated flow waveforms at simulated constant and pulse-modulated speed operations matched with those of the experimentally recorded flow waveforms. The simulations showed that the shear stress field and hemolysis index vary throughout the cardiac cycle under the constant speed operation, and also as a function of modulation profile under modulated speed operation. The computational model did not demonstrate any differences in the time average hemolysis index between constant and modulated pump speed operations, thereby predicting pulse-modulated speed operation may help to restore vascular pulsatility without any further increased risk of blood trauma. Graphical abstract The streamline inside the HVAD pump and the wall shear stress distribution on the impeller surface at six discrete time instants over one cardiac cycle under constant speed operation (3000 rpm) (a) and under pulse-modulated speed operation (b). c Computationally predicted flow rate waveform under pulse-modulated speed operation. d Computationally predicted time-varying HI generated by the HVAD pump under the two operation modes constant speed (dash line) and pulse-modulated speed (solid line). These figures indicate that the pulse-modulated speed operation may help to restore vascular pulsatility without any further increased risk of blood trauma.
More Related Videos
08:50Gene Expression Analysis of Endothelial Cells Exposed to Shear Stress Using Multiple Parallel-plate Flow Chambers
Published on: October 21, 2018
09:12Live Cell Analysis of Shear Stress on Pseudomonas aeruginosa Using an Automated Higher-Throughput Microfluidic System
Published on: January 16, 2019
Related Concept Videos
Shearing Stress
The average shearing stress can be calculated by dividing the shear by the area of the cross-section.
Shearing Stresses in a Beam: Problem Solving
Elastic Strain Energy for Shearing Stresses
Pulse
The pulse serves as a clinical...
Pulse
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls...
Constant Pressure Calorimetry