Comparison of meibomian gland dropout using two infrared imaging devices
Stephanie Wong1, Sruthi Srinivasan2, Paul J Murphy3
1Centre for Ocular Research & Education, School of Optometry and Vision Science, University of Waterloo, 200 University Avenue West, Waterloo, N2L 3G1, Canada.
Purpose:
To measure the degree of meibomian gland (MG) dropout in the lower eyelid determined by analysis of images obtained from the LipiView II (LVII) and the Keratograph 5M (K5M).
Methods:
The inferior eyelid of each participant was imaged in a random order using both devices. All images were subjectively assessed by a single-masked investigator to determine the extent of MG loss using the Arita 4-point meiboscore grading scale. The images were also semi-objectively analyzed with ImageJ to calculate the percentage of MG dropout, by tracing around the non-glandular area and the total exposed area of the lower lid.
Results:
Twenty participants (mean age 37 years, range 23-60, 60% female) completed the study. A significant difference in meiboscore (mean ± SD) was obtained between the LVII and the K5M (1.43 ± 0.78 vs. 1.90 ± 0.81, Z = 3.25, p = 0.001). The meiboscore 95% limit of agreement (LOA) ranged from -1.88 to +0.93. A significant difference was found with mean ImageJ percentage dropout between the LVII and the K5M (31.5% vs 43.4%, t = -4.8, p = 0.00003). The percentage dropout 95% LOA ranged from -42.79% to +19.06%.
Conclusions:
LVII images had significantly lower meiboscores and less percentage MG dropout. Varying amounts of dropout were observed between the devices due the amount of eyelid that was typically everted and because of differences in image quality. These results indicate that these devices should not be used interchangeably to evaluate MG dropout.
More Related Videos
09:33Using Near-Infrared Spectroscopy Wearable Devices to Identify Central Versus Peripheral Limitations During Exercise
Published on: December 19, 2024
13:10Combined Near-infrared Fluorescent Imaging and Micro-computed Tomography for Directly Visualizing Cerebral Thromboemboli
Published on: September 25, 2016
Related Concept Videos
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Exocrine Glands: Unicellular and Multicellular Glands
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
The Pituitary Gland
The Parathyroid Glands
Oxyphil cells, whose functions remain elusive, emerge during late puberty, adding a layer of complexity to the parathyroid gland's intricacies. In contrast, principal parathyroid cells undertake a vital role by...
The Pineal Gland
The primary secretion of the pineal gland is the hormone melatonin, derived from serotonin. The concentration of melatonin in the...