A novel calibration method incorporating nonlinear optimization and ball-bearing markers for cone-beam CT with a
Guang Li1,2, Shouhua Luo1, Chenyu You3
1Department of Biomedical Engineering, Southeast University, Nanjing, 210096, China.
Purpose:
Cone-beam (CB) CT is a powerful noninvasive imaging modality, and is widely used in many applications. Accurate geometric parameters are essential for high-quality image reconstruction. Usually, a CBCT system with higher spatial resolution, particularly on the order of microns or nanometers, will be more sensitive to the parametric accuracy. Here, we propose a novel calibration method combining a simple phantom containing ball bearing markers and an advanced optimization procedure. This method can be applied to CBCT with reproducible geometry and frame-to-frame invariant geometric parameters.
Methods:
Our proposed simplex-simulated annealing procedure minimizes the cost function that associates the geometrical parameters with the degree to which the back projections of the ball bearings in projections from various viewing angles converge, and the global minimum of the cost function corresponds to the actual geometric parameters. Specifically, six geometric parameters can be directly obtained by minimizing the cost function, and the last parameter, the distance from source to rotation axis (SRD), can be obtained using prior knowledge of the phantom - the spacing between the two ball bearings.
Results:
Numerical simulation was performed to validate that the proposed method with various noise levels. With the proposed method, the mean errors and standard deviations can be reduced to ∼10% and less than 1/3 of a competing benchmark method in the case of strong Gaussian noise (sigma = 200% of the pixel size) and large tilt angle (tilt angle = ). The calibration experiments with micro-CT and high-resolution CT scanners demonstrate that the proposed method recovers imaging parameters accurately, leading to superior image quality.
Conclusion:
The proposed method can obtain accurate geometric parameters of a CBCT system with a circular trajectory. While in the case of micro-CT the proposed method has a performance comparable to the competing method, for high-resolution CT, which is more sensitive to geometric calibration, the proposed method demonstrates higher calibration accuracy and more robustness than the benchmark algorithm.
More Related Videos
05:49Author Spotlight: Advancing CBCT and Digital Dental Image Integration with AI-Assisted Digitization
Published on: February 23, 2024
13:05Reliable Mechanochemistry: Protocols for Reproducible Outcomes of Neat and Liquid Assisted Ball-mill Grinding Experiments
Published on: January 23, 2018
Related Concept Videos
Orthogonal Trajectories
Nonlinear Pharmacokinetics: Causes of Nonlinearity
Nonlinear drug absorption can occur when the process is rate-limited by solubility, carrier-mediated transport systems, or saturation of the presystemic gut wall or hepatic metabolism. For instance, high doses of riboflavin...
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Collar Bearings