Artifact Properties of Dental Ceramic and Titanium Implants in MRI
Margit-Ann Geibel1, Benjamin Gelißen2, Anna-Katinka Bracher3
1Oral and Maxillofacial Surgery, Ulm-University, Ulm Germany.
Aim:
Assessment of the visualization of titanium and ceramic dental implants using various isotropic three-dimensional magnetic resonance imaging (MRI) methods.
Materials And Methods:
21 dental implants (7 ceramic, 14 titanium) were scanned in vitro with a spatially isotropically resolved three-dimensional gradient echo (FFE), a turbo spin echo (SE) and an ultra-short-echo time (UTE) imaging technique. The resulting absolute volumes of the implants were quantified and the relative error to the theoretical volume was calculated.
Results:
Ceramic implants and their periphery could be displayed well in all cases. The observed mean relative error results were 5.4 ± 2.3 % (UTE) to 6.5 ± 4.3 % (FFE). No significant difference was observed between the investigated MRI methods. The transition between implant and surrounding agarose could be shown in all cases without artifacts. Titanium implants resulted in mean relative errors between 1314 ± 350 % (FFE) and 2157 ± 810 % (SE). Here, significant differences were observed between the FFE and the SE and between the UTE and the SE sequence. The periphery of the implants could not be displayed in any case.
Conclusion:
Use of the MRI technique for the diagnosis of peri-implantitis, the assessment of anatomical structures and planning of dental implantation is currently very limited but could be used more frequently, provided there are no disturbing or imaging-disturbing materials in the region of interest. MRI technology is not suitable in case of titanium implants. When using ceramic implants, MRI technology is an option.
Key Points:
· MRI allows the artifact-free depiction of dental ceramic implants.. · Titanium implants cause the greatest relative errors in SE techniques.. · The UTE technique shows no significant improvements with respect to artifact behavior over the FFE technique..
Citation Format:
· Geibel M, Gelißen B, Bracher A et al. Artefakt-Verhalten von dentalen Keramik- und Titanimplantaten im MRT. Fortschr Röntgenstr 2019; 191: 433 - 441.
Related Concept Videos
Physical and Chemical Properties of Matter
Properties of Transition Metals
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
General Properties of Solutions
Convolution Properties I
The commutative property reveals that the input and the impulse response of an LTI (Linear Time-Invariant) system can be interchanged without affecting the output:
Properties of DTFT II
The frequency differentiation property is illustrated by considering a DTFT pair and differentiating both sides with respect to ω.