Diffusion-weighted MRI of the lung at 3T evaluated using echo-planar-based and single-shot turbo spin-echo-based
Neelam Tyagi1, Michelle Cloutier1, Kristen Zakian1
1Department of Medical Physics, Memorial Sloan Kettering Cancer Center, New York, NY, USA.
Purpose:
To compare single-shot echo-planar (SS-EPI)-based and turbo spin-echo (SS-TSE)-based diffusion-weighted imaging (DWI) in Non-Small Cell Lung Cancer (NSCLC) patients and to characterize the distributions of apparent diffusion coefficient (ADC) values generated by the two techniques.
Methods:
Ten NSCLC patients were enrolled in a prospective IRB-approved study to compare and optimize DWI using EPI and TSE-based techniques for radiotherapy planning. The imaging protocol included axial T2w, EPI-based DWI and TSE-based DWI on a 3 T Philips scanner. Both EPI-based and TSE-based DWI sequences used three b values (0, 400, and 800 s/mm2 ). The acquisition times for EPI-based and TSE-based DWI were 5 and 8 min, respectively. DW-MR images were manually coregistered with axial T2w images, and tumor volume contoured on T2w images were mapped onto the DWI scans. A pixel-by-pixel fit of tumor ADC was calculated based on monoexponential signal behavior. Tumor ADC mean, standard deviation, kurtosis, and skewness were calculated and compared between EPI and TSE-based DWI. Image distortion and ADC values between the two techniques were also quantified using fieldmap analysis and a NIST traceable ice-water diffusion phantom, respectively.
Results:
The mean ADC for EPI and TSE-based DWI were 1.282 ± 0.42 × 10-3 and 1.211 ± 0.31 × 10-3 mm2 /s. The average skewness and kurtosis were 0.14 ± 0.4 and 2.43 ± 0.40 for DWI-EPI and -0.06 ± 0.69 and 2.89 ± 0.62 for DWI-TSE. Fieldmap analysis showed a mean distortion of 13.72 ± 8.12 mm for GTV for DWI-EPI and 0.61 ± 0.4 mm for DWI-TSE. ADC values obtained using the diffusion phantom for the two techniques were within 0.03 × 10-3 mm2 /s with respect to each other as well as the established values.
Conclusions:
Diffusion-weighted turbo spin-echo shows better geometrical accuracy compared to DWI-EPI. Mean ADC values were similar with both acquisitions but the shape of the histograms was different based on the skewness and kurtosis values. The impact of differences in respiratory technique on ADC values requires further investigation.
Related Concept Videos
Echo
Imagine the sound is reflected back to the ears. Assuming that the source is very close to the human, the difference between hearing the two sounds—the emitted sound and the reflected sound—may be more than the minimum time for perceiving distinct sounds. If this is the case,...
Crystal Field Theory - Tetrahedral and Square Planar Complexes
Crystal field theory (CFT) is applicable to molecules in geometries other than octahedral. In octahedral complexes, the lobes of the dx2−y2 and dz2 orbitals point directly at the ligands. For tetrahedral complexes, the d orbitals remain in place, but with only four ligands located between the axes. None of the orbitals points directly at the tetrahedral ligands. However, the dx2−y2 and dz2 orbitals (along the Cartesian axes) overlap with the ligands less than the dxy,...
Diffusion
DNA Base Pairing
Lewis Acids and Bases
A coordinate covalent bond (or dative bond) occurs when one of the atoms in the bond provides both bonding electrons. For example, a coordinate covalent bond occurs when a water molecule combines with a hydrogen ion to form a hydronium ion. A coordinate covalent bond also results when...
Base Excision Repair
The first step of...