Plasma lactate measurement as an example of encountered gaps between routine clinical laboratory processes and
Ibrahim A Hashim1,2, Mishkat Mohamed1, Aileen Cox1
1Clinical Laboratories, Parkland Health and Hospital System, United States.
Objectives:
Deviation from manufacturers' pre-analytical sample handling recommendations necessitates extensive validation studies. This report uses plasma lactate testing, where a recommended 15 min room temperature sample handling limit cannot be met by the clinical laboratory, as an example for studies to bridge the gap with practice.
Design And Methods:
Triplicate blood samples were collected from patients (n = 51) with lactate requests by clinicians and from normal volunteers (n = 50). One tube was transported on ice (4 °C), the others were maintained at room temperature (23 °C). Tubes stored at 4 °C were processed at 30 min from collection. Tubes stored at 23 °C were processed at 15 and at 30 min from collection. Lactate levels were measured using Roche Diagnostics Cobas 6000® analyzer.
Results:
Lactate levels in normal subjects ranged from 0.6 to 3.1 mmol/L (median 1.1). Patient lactate levels ranged from 0.8 to 26.3 mmol/L (median 2.2). Bias in lactate levels following extended storage of samples from both normal subjects and patients ranged from - 1.3 to2.2 and from - 1.0-1.0 mmol/L when stored for 30 min at 23 °C or at 4 °C, respectively. The bias between lactate levels at 30 min at 23 °C and 4 °C was - 1.2 to - 0.5 mmol/L for both populations. Although the bias was not statistically significant for all variables, a clinically significant (>0.2 mmol/L) bias was observed in 28% of normal and 7.0% of patient samples.
Conclusion:
Extending the pre-analytical time to 30 min at 23 °C did not significantly impact clinical utility of lactate measurement in our patient population.