Related Experiment Video
Updated: Feb 2, 2026

Evaluation of Nanoparticle Uptake in Tumors in Real Time Using Intravital Imaging
Published on: June 21, 2011
Evaluation of a real-time low-power cardiorespiratory sensor for the IoT
Abstract:
A wide variety of sensors have been developed in the biomedical engineering community for telemedicine and personalized healthcare applications. However, they usually focus on sensor connectivity and embedded signal processing, at the expense of the sensing part. This observation lead to the development and exhaustive evaluation of a new ECGbased cardiorespiratory IoT sensor. In order to improve the robustness of our IoT-based sensor, we discuss in detail the influence of electrodes placement and nature. Performance assessment of our sensor resulted in a best-case sensitivity of 99.95% and a precision of 99.89% for an abdominal positioning of wet electrodes, while a sensitivity of 99.47% and a precision of 99.31% were observed using a commercialgrade dry electrodes belt. Consequently, we prove that our sensor is fit for the comfortable medical-grade monitoring of the cardiorespiratory activity in order to provide insights of patients health in a telemedicine context.