Related Experiment Video
Updated: Feb 2, 2026

Determination of Inorganic Arsenic in a Wide Range of Food Matrices using Hydride Generation - Atomic Absorption Spectrometry.
Published on: September 1, 2017
Solid Phase Extraction of Cadmium and Lead from Water by Amberlyst 15 and Determination by Flame Atomic Absorption
Adalet Tunçeli1, Abdullah Ulaş1, Orhan Acar1
1Department of Chemistry, Science Faculty, Gazi University, 06500, Ankara, Turkey.
Abstract:
Preconcentration of Cd(II) and Pb(II) was carried out by using column solid phase extraction method. Amberlyst 15 was used as solid phase for these analytes. The optimum extraction conditions such as pH (4), type and volume of eluent (5 mL of 2 mol L-1 HNO3) sample flow rate (1 mL min-1) and sample volume (100 mL for Cd(II) and 750 mL for Pb(II)) were determined. The recoveries were found for Cd(II) and Pb(II) as 104% ± 1% and 102 % ± 2%, respectively. The limit of detections were found as 0.23 µg L-1 for Cd(II) and 0.13 µg L-1 for Pb(II). The effects of foreign ions were also studied. The method was validated by analyzing standard reference material and spiked water samples. Percent relative error and relative standard deviation were below 3% and 4%, respectively.
More Related Videos
Related Concept Videos
Phase-lead and Phase-lag Controllers
Atomic Absorption Spectroscopy: Atomization Methods
Mineral, Vitamin and Water Absorption
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Atomic Absorption Spectroscopy: Overview
When irradiated by EMR of a particular wavelength, these...
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...