Related Experiment Video
Updated: Feb 2, 2026

Determining Glucose Metabolism Kinetics Using 18F-FDG Micro-PET/CT
Published on: May 2, 2017
A new methodology to derive 3D kinetic parametric FDG PET images based on a mathematical approach integrating an
Elyse Colard1, Sarkis Delcourt2, Laetitia Padovani3
1LITIS-QuantIF-EA4108, University of Rouen, Rouen, France.
Background:
In FDG-PET, SUV images are hampered by large potential biases. Our aim was to develop an alternative method (ParaPET) to generate 3D kinetic parametric FDG-PET images easy to perform in clinical oncology.
Methods:
The key points of our method are the use of a new error model of PET measurement extracted from a late dynamic PET acquisition of 15 min, centered over the lesion and an image-derived input function (IDIF). The 15-min acquisition is reconstructed to obtain five images of FDG mean activity concentration and images of its variance to model errors of PET measurement. Our approach is carried out on each voxel to derive 3D kinetic parameter images. ParaPET was evaluated and compared to Patlak analysis as a reference. Hunter and Barbolosi methods (Barbolosi-Bl: with blood samples or Barbolosi-Im: with IDIF) were also investigated and compared to Patlak. Our evaluation was carried on Ki index, the net influx rate and its maximum value in the lesion (Ki,max).
Results:
This parameter was obtained from 41 non-small cell lung cancer lesions associated with 4 to 5 blood samples per patient, required for the Patlak analysis. Compare to Patlak, the median relative difference and associated range (median; [min;max]) in Ki,max estimates were not statistically significant (Wilcoxon test) for ParaPET (- 3.0%; [- 31.9%; 47.3%]; p = 0.08) but statistically significant for Barbolosi-Bl (- 8.0%; [- 30.8%; 53.7%]; p = 0.001), Barbolosi-Im (- 7.9%; [- 38.4%; 30.6%]; p = 0.007) or Hunter (32.8%; [- 14.6%; 132.2%]; p < 10- 5). In the Bland-Altman plots, the ratios between the four methods and Patlak are not dependent of the Ki magnitude, except for Hunter. The 95% limits of agreement are comparable for ParaPET (34.7%), Barbolosi-Bl (30.1%) and Barbolosi-Im (30.8%), lower to Hunter (81.1%). In the 25 lesions imaged before and during the radio-chemotherapy, the decrease in the FDG uptake (ΔSUVmax or ΔKi,max) is statistically more important (p < 0.02, Wilcoxon one-tailed test) when estimated from the Ki images than from the SUV images (additional median variation of - 2.3% [- 52.6%; + 19.1%] for ΔKi,max compared to ΔSUVmax).
Conclusion:
None of the four methodologies is yet ready to replace the Patlak approach, and further improvements are still required. Nevertheless, ParaPET remains a promising approach, offering a non-invasive alternative to methods based on multiple blood samples and only requiring a late PET acquisition. It allows deriving Ki values, highly correlated and presenting the lowest relative bias with Patlak estimates, in comparison to the other methods we evaluated. Moreover, ParaPET gives access to quantitative information at the pixel level, which needs to be evaluated in the perspective of radiomic and tumour response.
Trial Registration:
NCT 02821936 ; May 2016.
Related Concept Videos
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...
Mathematical Modeling: Problem Solving
Measurement: Derived Units
The Quantum-Mechanical Model of an Atom
Kinetic Molecular Theory: Molecular Velocities, Temperature, and Kinetic Energy
Fundamental Attribution Error

![Visualization and Quantification of Brown and Beige Adipose Tissues in Mice using [18F]FDG Micro-PET/MR Imaging](/_next/image?url=https%3A%2F%2Fcloudfront.jove.com%2FCDNSource%2Fteasers%2F62460.jpg&w=3840&q=50)