Effects of cone combinations on accurate and precise Mg-isotopic determination using multi-collector inductively
Long-Fei Gou1,2,3, Zhangdong Jin1,4, Albert Galy2
1State Key Laboratory of Loess and Quaternary Geology, Institute of Earth Environment, Chinese Academy of Sciences, Xi'an, 710061, China.
Rationale:
High-precision determination of magnesium (Mg) isotopes can now be routinely achieved by multi-collector inductively coupled plasma mass spectrometry (MC-ICP-MS). The analytical sensitivity and instrumental mass discrimination behavior of this method are, however, sensitive to the types of sample and skimmer cones used in these measurements, so it is important that these parameters should be investigated.
Methods:
Using the sample-standard-bracketing method in the wet-plasma mode, four available combinations of sample and skimmer cones [Jet sample cone + H skimmer cone (Jet + H), standard sample cone + H skimmer cone (Standard + H), standard sample cone + X skimmer cone (Standard + X), and Jet sample cone + X skimmer cone (Jet + X)] were systematically investigated for peak shape, sensitivity, mass discrimination, accuracy, and precision in Mg-isotopic ratio determination using a Neptune plus MC-ICP mass spectrometer.
Results:
The results showed that different cone combinations do not affect peak shapes but would significantly change the sensitivities for Mg-isotopic determinations. Compared with using the Standard + H, the sensitivities of Mg-isotopic determinations were enhanced by approximately a factor of 1.3, 1.4, and 1.9 by using the Standard + X, the Jet + H, and the Jet + X combinations, with the most stable mass discrimination behaviors obtained by the Jet + H. The instrumental mass fractionation slope for any combination of a modified cone geometry (i.e. Standard + X, Jet + X, and Jet + H) is 0.500, while it is 0.510 for the Standard + H. In addition, the mass discrimination behavior is related to Mg concentrations once the combination is set, indicating the necessity of concentration match during Mg-isotopic determination.
Conclusions:
The precision and accuracy of the Jet + H combination are better than those of the other combinations, and this is further supported by the validation of the Mg-isotope data for four international reference materials: Cambridge-1, NASS-6, AGV-2, and BHVO-2. As the Jet + H combination also provides a high signal, this combination gives the most robust strategy for the highly precise and accurate determination of Mg isotopes.
More Related Videos
Related Concept Videos
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Mass Spectrometry: Isotope Effect
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
Isotopes
An element's atomic mass, or weight,...