Preventing condensation of objective lens in noncontact wide-angle viewing systems during vitrectomy
Jung Pil Lee1, Jinsoo Kim2, Inwon Park2
1Department of Ophthalmology, Incheon Medical Center, Incheon 22532, Republic of Korea.
Aim:
To assess the optimal conditions for preventing condensation of objective lens during vitrectomy with noncontact wide-angle viewing systems (WAVSs).
Methods:
We explored the effectiveness of the coating with ophthalmic viscoelastic device (OVDs) on the corneal surface and the soaking the objective lens in warm-saline for preventing condensation of objective lens. First, to find the optimal soaking time to keep the objective lens warm, we measured the temperature of objective lens every minute after soaking in warm saline. Second, to find optimal distance between cornea and objective lens, which provide as wide a view as possible and less condensation at the same time, we measured the condensation time with different distances. With the obtained optimal soaking time and distance, we explored the effect of coating cornea with OVDs and soaking objective lens in warm saline on condensation time.
Results:
One and 5min of soaking in warm saline was most effective for keeping the lens warm enough (45.1°C±2.1°C for 1min and 46.4°C ±1.0°C for 5min, P=0.109). The mean condensation times for the control group at 1, 3, and 5 mm from corneal surface to objective lens were 1±0.4, 4±1.4, 190±26.1s, respectively, thus 5 mm was most optimal distance for vitrectomy with WAVSs. For the OVD coating group, the mean condensation times were 1.5±0.3, 13±1.4, and 200±23.9s at 1, 3, and 5 mm distance and borderline significant compared with control group (P=0.068, 0.051, and 0.063, respectively). With the 1-minute warm saline soaking group, the mean condensation time were extended to 188±34.4, 416±65.7, and 600±121.3s at 1, 3, and 5 mm distance and statistically significant compared with control (P=0.043, 0.041 and 0.043, respectively).
Conclusion:
OVD coating on corneal surface shows no difference on condensation time with control group. However, soaking the objective lens in warm saline revealed statistically significant extension of condensation time compared to control group. Therefore, keeping the objective lens warm with soaking in warm saline is a simple but effective to prevent condensation of objective lens during vitrectomy. The thermodynamics between objective lens and cornea during vitrectomy warrants further investigation.
More Related Videos
11:27Studying Soft-matter and Biological Systems over a Wide Length-scale from Nanometer and Micrometer Sizes at the Small-angle Neutron Diffractometer KWS-2
Published on: December 8, 2016
07:11Author Spotlight: Insights into Visual Cortex Research Through Wide-View fMRI Mapping
Published on: December 8, 2023
Related Concept Videos
Aldol Condensation vs Claisen Condensation
Phase Transitions: Vaporization and Condensation
Aldol Condensation with β-Diesters: Knoevenagel Condensation
Contact Angle
The adhesive force is the molecular force between molecules of different materials, that is, between the molecules of the solid and the liquid. The cohesive...
Econometric Views (EViews)
Velocity of an Object