Simultaneous multi-slice cardiac cine with Fourier-encoded self-calibration at 7 Tesla
Stanislas Rapacchi1,2, Thomas Troalen3, Zakarya Bentatou1
1Aix Marseille Univ, CNRS, CRMBM, Marseille, France.
Purpose:
To accelerate cardiac cine at 7 tesla using simultaneous multi-slice (SMS) acquisition with self-calibration to resolve misalignment between calibration and imaging data due to breathing motion.
Methods:
A spoiled-gradient echo cine sequence was modified with radiofrequency phase-cycled SMS excitations. A Fourier encoding strategy was applied along the cardiac phase dimension to allow for slice untangling and split-slice GRAPPA calibration. Split-slice GRAPPA was coupled with regular GRAPPA (SMS-GRAPPA) and L1-SPIRiT (SMS-L1SPIRiT) for image reconstruction. 3-slice SMS cine MRI was evaluated in ten subjects against single-slice cine MRI in terms of SNR and contrast-to-noise ratio and slice leakage.
Results:
SNR decreased significantly from 10.1 ± 7.1 for single-slice cine to 7.4 ± 2.8 for SMS-GRAPPA (P = 0.02) and was recovered to 9.0 ± 4.5 with SMS-L1SPIRiT (P = 0.02). Contrast to noise ratio decreased significantly from 14.5 ± 8.1 for single-slice cine to 5.6 ± 3.6 for SMS-GRAPPA (P < 0.0001) and increased slightly but significantly back to 6.7 ± 4.4 for SMS-L1SPIRiT (P = 0.03). Specific absorption rate restrictions imposed a reduced nominal flip angle (-37 ± 7%, P = 0.02) for 3-slice SMS excitations compared to single-slice acquisitions. SMS slice leakage increased significantly from apex (8.6 ± 6.5 %) to base (13.1 ± 4.1 %, P = 0.03) in the left ventricle.
Conclusion:
Three-fold acceleration of cine at 7T was achieved using the proposed SMS technique. Fourier encoding self-calibration and regularized image reconstruction enabled simultaneous acquisition of three slices without significant SNR decrease but significant CNR decrease linked to the reduced nominal excitation flip angle.
Related Concept Videos
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Encoding
Automatic processing involves the encoding of details like time, space, frequency, and the meaning of words, usually done without conscious...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Trigonometric Fourier series
The trigonometric Fourier series specifically expresses a periodic function with a defined period T using sine...
Convergence of Fourier Series
The Gibbs phenomenon refers to the persistent oscillations and overshoots that occur near discontinuities...
Fast Fourier Transform
The computational efficiency of the FFT becomes...