Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
Accuracy of a Smartphone-based Autorefractor Compared with Criterion-standard Refraction
V Swetha E Jeganathan1, Nita Valikodath1, Leslie M Niziol1
1Department of Ophthalmology and Visual Sciences, W. K. Kellogg Eye Center, University of Michigan, Ann Arbor, Michigan.
Significance:
Uncorrected refractive error is a prevalent problem throughout the world especially among the low-income population who have limited access to professional eye care and cannot afford eyeglasses.
Purpose:
The purpose of this study was to evaluate the accuracy and usability of a low-cost, portable, smartphone-based autorefractor (Netra, EyeNetra Inc., Somerville, MA) in adults.
Methods:
A cross-sectional study was conducted to compare the portable refractor with subjective (manifest and cycloplegic) refraction for sequential adult participants with best-corrected visual acuity of 20/40 or greater. For each method of refraction, the spherical equivalent was calculated. Differences between methods were tested with linear mixed regression models. A validated usability questionnaire was administered regarding ease of use (100-point scale, higher scores better) for the portable autorefractor.
Results:
Eighty-seven subjects (152 eyes) were studied (age range, 20 to 90 years; mean ± standard deviation, 51.9 ± 18.3 years). Mean spherical equivalent by the portable device was -2.76 D (range, -14.75 to 3.63 D) compared with -2.49 D (range, -15.25 to 4.25 D) by manifest refraction. The mean relative difference in spherical equivalent between methods was -0.27 D (P = .001, significantly different than 0 D). The mean absolute difference between methods was 0.69 D (P < .001, significantly different than 0.5-D absolute difference). Similar results were found when comparing spherical equivalent between Netra and cycloplegic refraction methods. Subjects reported average ease of use for the Netra of 75.4 ± 19.8.
Conclusions:
The portable autorefractor had small but clinically significant differences from subjective refraction. The device's scores on the usability scale indicate good overall patient acceptance. The device may be valuable for use where there is limited access to a trained refractionist.
Related Concept Videos
Improving Translational Accuracy
Uncertainty in Measurement: Accuracy and Precision
Routh-Hurwitz Criterion I
To apply the Routh-Hurwitz criterion, a Routh table is constructed. The table's rows are labeled with powers of the complex frequency variable s, starting from the...
Routh-Hurwitz Criterion II
The first scenario occurs when a singular zero appears in the first column of the Routh table. This situation creates a division by zero issues. To resolve this, a small positive or negative number, denoted as epsilon (∈), is substituted for the zero. The stability analysis proceeds by assuming a sign for ∈. If ∈ is positive, any sign change in the first...
Pharmacokinetic Models: Comparison and Selection Criterion
Physiological models take a detailed approach by considering specific molecular processes. They can predict drug distribution, metabolism, and elimination changes, providing a comprehensive understanding of how drugs interact with the body.
Potential-Energy Criterion for Equilibrium