Comparison of the Ultraviolet Light Filtering across Different Intraocular Lenses
Mari Carmen García-Domene1, Caridad Pérez-Vives2,3, Cristina Peris-Martínez1
1FISABIO-Oftalmología Médica, Valencia, Spain.
Significance:
We have analyzed the ultraviolet transmittance of some commercial intraocular lenses (IOLs). The results show differences of wavelength cutoff among them.
Purpose:
The purpose of this study was to measure and compare the ultraviolet light transmittance of different IOLs made out of acrylic hydrophobic, hydrophilic, and hydrophilic with hydrophobic surface materials from different manufacturers.
Methods:
The spectral transmission curves of eight monofocal IOLs with the same dioptric power of +20.0 diopters were measured using a PerkinElmer Lambda 35 ultraviolet/visible spectrometer. Two IOLs of each type were tested three times. The ultraviolet cutoff wavelength at 10% transmission and the mean values were calculated.
Results:
All lenses prevented transmission of ultraviolet C (200 to 280 nm) and B radiation (280 to 315 nm). However, not all IOLs provided the same filtering properties in ultraviolet A (315 to 380 nm). Within the ultraviolet A range, the ultraviolet radiation cutoff wavelength of 10% ranges from approximately 360 to 400 nm. HOYA iSert 250 provided a cutoff wavelength of 398.4 nm; AcrySof SA60AT, 396.2 nm; AcrySof SA60WF, 395.7 nm; CT Asphina 404, 378.34 nm; Tecnis ZCB00, 377.70 nm; CT Lucia 607P, 379 nm; C-Flex 570C, 377 nm; and enVista MX60, 360 nm.
Conclusions:
Intraocular lenses of different materials and manufacturers have different ultraviolet transmission characteristics. AcrySof (SA60AT and SA60WF) and HOYA iSert 250 provided the highest ultraviolet radiation transmission; the cutoff wavelength of 10% is close to 400 nm. In contrast, enVista IOL showed the lowest ultraviolet radiation cutoff.
Related Concept Videos
Passive Filters
Low-Pass Filters
Low-pass filters are designed to transmit signals with frequencies lower than the cutoff frequency, ωc, and attenuate those above it. The cutoff...
Active Filters
Photoreceptors and Plant Responses to Light
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Ultraviolet and Visible (UV–Vis) Spectroscopy: Overview
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...