ESTIMATION OF PATIENT LENS DOSE ASSOCIATED WITH C-ARM CONE-BEAM COMPUTED TOMOGRAPHY USAGE DURING INTERVENTIONAL
Satoru Kawauchi1,2, Koichi Chida2, Takashi Moritake3
1Department of Radiology, Toranomon Hospital, 2-2-2 Toranomon, Minato, Tokyo, Japan.
Abstract:
The purpose of this study was to investigate the dose distribution and lens doses associated with C-arm cone-beam computed tomography (CBCT), using a head phantom, and to estimate the contribution ratio of C-arm CBCT to each patient's lens dose during interventional neuroradiology ('lens dose ratio') in 109 clinical cases. In the phantom study, the peak skin doses and respective right and left lens doses of C-arm CBCT were as follows: 63.0 ± 1.9 mGy, 19.7 ± 1.4 mGy and 21.9 ± 0.8 mGy in whole brain C-arm CBCT and 39.2 ± 1.4 mGy, 4.7 ± 0.9 mGy and 3.6 ± 0.3 mGy in high-resolution C-arm CBCT. In the clinical study, the lens dose ratios were 25.4 ± 8.7% in the right lens and 19.1 ± 9.8% in the left lens. This study shows that, on average, ~25% of patients' total lens dose was contributed by C-arm CBCT.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Drug Dosing: Geriatric Patients
Drug Dosing: Obese Patients
Imaging Studies III: Computed Tomography
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
Muscles that Move the Arm
The pectoralis major has two origins. Its clavicular head originates on the medial half of the clavicle. In contrast, the sternocostal head originates on the costal cartilages of ribs 1-6, the sternum, and the aponeurosis of the external oblique of the...