Dependence of volume dose indices on dose calculation algorithms for VMAT-SBRT plans for peripheral lung tumor
Takanori Matsuoka1, Fujio Araki2, Takeshi Ohno2
1Graduate School of Health Sciences, Kumamoto University, Kumamoto, Japan.
Abstract:
The purpose of this study was to investigate the dependence of volume dose indices on dose calculation algorithms for volumetric modulated arc therapy (VMAT) for stereotactic body radiotherapy (SBRT) plans to treat peripheral lung tumors by comparing them with those of Monte Carlo (MC) calculations. VMAT-SBRT plans for peripheral lung tumors were created using the Eclipse treatment planning system (TPS) for 24 patients with nonsmall cell lung cancer. VMAT dose distributions for gross tumor volume (GTV), internal target volume (ITV), and planning target volume (PTV) were calculated using the analytical anisotropic algorithm (AAA), the Acuros XB (AXB) algorithm, and a MC algorithm. VMAT dose distributions of the 3 algorithms were compared using their volume dose indices from dose volume histograms (DVHs), a dose difference map, and 3-dimensional gamma analysis. The DVHs for GTV and ITV from AAA, AXB, and MC were in good agreement. The difference between the ITV and PTV volume dose indices from AAA and MC increased as D98, D95, D80, D50, and D2. In particular, the difference between D98 for PTV from AAA and MC was up to 48%. A >5% difference between D95 for PTV from AAA and MC was 11 patients, but only 2 patients for ITV. The volume dose indices for AXB were near those of MC. AAA tended to overestimate the PTV volume dose indices compared to AXB and MC. Thus, we propose that the volume dose indices for the ITV be used because they are independent of dose calculation algorithms.
Related Concept Videos
Nonlinear Pharmacokinetics: Dependence of Elimination Half-Life and Dose Clearance
A study on guinea pigs examined the...
Dose Size and Dosing Frequency: Determination Methods
Determination of Multiple Dosing Parameters: Loading and Maintenance Doses
Rational Dosage Regimen: Maintenance Dose and Loading Dose
In most cases, drugs are administered repetitively or infused continuously to maintain a steady-state concentration in the body. At a steady...
Dose-Response Relationship: Overview
Drug Dosing: Infants and Children