Validation of Biomode S.A. Probe4CronobacterTM for the Identification of Cronobacter spp
J Mário Sousa1, Rui Rocha1, Laura Cerqueira1
1Biomode 2, S.A., Edifício GNRation, Praça Conde Agrologo nº123, Braga 4700-312, Portugal.
Abstract:
Background: Probe4Cronobacter test kit is based on the use of a fluorescence-labeled peptide nucleic acid probe (PNA) allied to fluorescence microscopy. A sample is taken after a 24 h enrichment of rehydrated 30 g portions of powdered infant formula (PIF). The method uses ready to use dropper solutions applied directly in the sample. This simple process takes less than 2 h to provide a result. In the presence of Cronobacter species, bright red rod-shaped cells will be visible under a fluorescence microscope. Objective: Probe4Cronobacter validation as a new method for the detection Cronobacter species in Powdered Infant Formula (PIF) under the AOAC Performance Tested MethodsSM (License No. 081702). Methods: The validation study encompassed matrix comparison study, inclusivity and exclusivity testing and robustness studies (stability, kit variation, and ruggedness). Results: The inclusivity and exclusivity testing (50 and 35 strains, respectively) yielded no false negative or false positive results. Probe4Cronobacter was compared to the ISO/TS 22964:2006 in 30 g of PIF samples within method comparison in an unpaired study. A total of 30 samples with both low and high level of inoculation were analyzed by Probe4Cronobacter and compared to the same number of samples screened by ISO/TS 22964:2006. No statistically significant differences between presumptive and confirmed results or between candidate and reference method results were observed. Robustness studies showed a high level of consistency and integrity of the kit when different parameters were varied. The deviation conditions tested did not affect the performance of the kit. Conclusions: Probe4Cronobacter test kit has shown to be a accurate, highly sensitive and robust methods for the detection of Cronobacter spp. in PIF samples.
Related Concept Videos
Reliability and Validity
Data Validation
Key parameters for method validation include:
Data Validation
Nursing assessment guides are generally based on holistic models rather than medical...
Formulating and Validating Nursing Diagnosis I
There are thirteen domains...
Formulating and Validating Nursing Diagnosis II
Risk nursing diagnoses represent clinical judgments of an individual, family, or community more vulnerable to developing the health problem than others...
Methods of Classification and Identification