Radio-Frequency Safety Assessment of Stents in Blood Vessels During Magnetic Resonance Imaging
Kyoko Fujimoto1, Leonardo M Angelone1, Elena Lucano1
1Division of Biomedical Physics, Office of Science and Engineering Laboratory, Center for Devices and Radiological Health, U.S. Food and Drug Administration, Silver Spring, MD, United States.
Abstract:
Purpose: The purpose of this study was to investigate the need for high-resolution detailed anatomical modeling to correctly estimate radio-frequency (RF) safety during magnetic resonance imaging (MRI). RF-induced heating near metallic implanted devices depends on the electric field tangential to the device (E ). E and specific absorption rate (SAR) were analyzed in blood vessels of an anatomical model to understand if a standard gel phantom accurately represents the potential heating in tissues due to passive vascular implants such as stents. Methods: A numerical model of an RF birdcage body coil and an anatomically realistic virtual patient with a native spatial resolution of 1 mm3 were used to simulate the in vivo electric field at 64 MHz (1.5 T MRI system). Maximum values of SAR inside the blood vessels were calculated and compared with peaks in a numerical model of the ASTM gel phantom to see if the results from the simplified and homogeneous gel phantom were comparable to the results from the anatomical model. E values were also calculated in selected stent trajectories inside blood vessels and compared with the ASTM result. Results: Peak SAR values in blood vessels were up to ten times higher than those found in the ASTM standard gel phantom. Peaks were found in clinically significant anatomical locations, where stents are implanted as per intended use. Furthermore, E results showed that volume-averaged SAR values might not be sufficient to assess RF safety. Conclusion: Computational modeling with a high-resolution anatomical model indicated higher values of the incident electric field compared to the standard testing approach. Further investigation will help develop a robust safety testing method which reflects clinically realistic conditions.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies IV: Magnetic Resonance Imaging
Structure of Blood Vessels
Anatomy of Blood Vessels
Arteries
Arteries circulate oxygenated blood from the heart, except the pulmonary artery, which transports deoxygenated blood to the lungs. Large arteries, such as the aorta,...
Development of Blood Vessels
The initial formation of this system is facilitated by the small amount of yolk present in the ovum and yolk sac. Blood vessels originate from...
Overview of Blood Vessels
Arteries and Arterioles: Arteries are muscular and elastic vessels that primarily carry oxygenated blood from the heart to body tissues, except for the pulmonary artery, which carries deoxygenated blood. They have thick walls to withstand high pressure and contain a layer of muscle tissue, allowing them to expand or contract as...