Evaluation of Probe Angles for Synchronous Waveform 915-MHz Microwave Ablation
Yan Li1, Megan Stevenson1, Jack W Rostas1
1Division of Surgical Oncology, Department of Surgery, University of Louisville, Louisville, Kentucky.
Background:
The purpose of this study is to compare the ablation performance between a synchronous microwave ablation (MWA) system and a commercially available asynchronous system in ex vivo bovine liver and evaluate the efficacy of ablation at varying entrance angles.
Materials And Methods:
Two 915-MHz MWA systems were used in bench top ex vivo bovine livers with various conditions (synchronous versus asynchronous). Using synchronous technology ablations to liver, kidney, or lung at angles of 0, 15, 30, and 90° were evaluated.
Results:
Synchronous and asynchronous MWA systems created mean ablation zone volumes of 26.4 and 15.8 cm3, 62.9 and 45.4 cm3, 90.8 and 56.4, and 75.7 and 54.8 cm3 with single, double (2 microwave probes in use simultaneous) (2 cm spacing), and triple (three probes in use simultaneously) (2 cm and 3 cm spacing) antennae configurations, respectively; adjusted P-values ≤ 0.006. Ablation defects were similar across all groups when evaluated for entrance angle. Specifically, when comparing 0-degree angle to all other angles, achieved zones of ablation (ZA) were similar (mean ± standard deviation for 0-degree versus all other angles: 8.72 ± 4.84 versus 9.38 ± 4.11 cm2, P = 0.75). The use of the long-tip probe resulted in a statistically significant increase in the achieved ZA when compared to the short tip probe (10.9 ± 4.3 versus 6.5 ± 2.4, respectively; P = 0.01).
Conclusions:
Newly developed synchronous microwave technology creates significantly larger ablation zones when compared to an existing asynchronous commercially available system. The angle of approach does not affect the resulting ZA. This is clinically relevant as true 0-degree angle is often difficult to obtain.
Related Concept Videos
Effective Value of a Periodic Waveform
The effective value of a periodic current represents the direct current (DC) that conveys the same average power to a resistor as the periodic current itself. This concept is crucial when assessing AC circuits. To determine the...
Contact Angle
The adhesive force is the molecular force between molecules of different materials, that is, between the molecules of the solid and the liquid. The cohesive...
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Simplified Synchronous Machine Model
In this model, each generator is connected to a...
Angle of Twist: Problem Solving
Angle Closure Glaucoma: Treatment