Objective:
To evaluate a prototype automated controller (IntellO2) of the inspired fraction of oxygen (FiO2) in maintaining a target range of oxygen saturation (SpO2) in preterm babies receiving nasal high flow (HF) via the Vapotherm Precision Flow.
Design:
Prospective two-centre order-randomised cross-over study.
Setting:
Neonatal intensive care units.
Patients:
Preterm infants receiving HF with FiO2 ≥25%.
Intervention:
Automated versus manual control of FiO2 to maintain a target SpO2 range of 90%-95% (or 90%-100% if FiO2=21%).
Main Outcome Measures:
The primary outcome measure was per cent of time spent within target SpO2 range. Secondary outcomes included the overall proportion and durations of SpO2 within specified hyperoxic and hypoxic ranges and the number of in-range episodes per hour.
Results:
Data were analysed from 30 preterm infants with median (IQR) gestation at birth of 26 (24-27) weeks, study age of 29 (18-53) days and study weight 1080 (959-1443) g. The target SpO2 range was achieved 80% of the time on automated (IntellO2) control (IQR 70%-87%) compared with 49% under manual control (IQR 40%-57%; p<0.0001). There were fewer episodes of SpO2 below 80% lasting at least 60 s under automated control (0 (IQR 0-1.25)) compared with manual control (5 (IQR 2.75-14)). There were no differences in the number of episodes per hour of SpO2 above 98% (4.5 (IQR 1.8-8.5) vs 5.5 (IQR 1.9-14); p=0.572) between the study arms.
Conclusions:
The IntellO2 automated oxygen controller maintained patients in the target SpO2 range significantly better than manual adjustments in preterm babies receiving HF.
Trial Registration Number:
NCT02074774.
More Related Videos
Related Concept Videos
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Control of Power Flow
Crossing Over
The homologous pairs of sister chromosomes—one from the maternal and one from the paternal genome—then begin to align alongside each other lengthwise, matching corresponding DNA positions in a process...
Types of Biopharmaceutical Studies: Controlled and Non-Controlled Approaches
Non-controlled studies, commonly employed for initial exploration, lack a control group, rendering them susceptible to biases and external influences. In contrast,...
Cross-Sectional Research