[Imaging Parameter Optimization of 3D Phase Contrast-MRA to Reduce Susceptibility-artifact and
Masato Kiriki1, Wataru Jomoto1, Takashi Ikeda1
1Department of Radiological Technology, Hyogo College of Medicine College Hospital.
Abstract:
The purpose of this study was to evaluate the degree of signal attenuation generated around the stent by the susceptibility artifacts and radiofrequency (RF) shielding in phase contrast-magnetic resonance angiography (PCA) images and construct optimal PCA imaging parameters for intracranial stent lumen images. The two types of PCA images of stents (Neuroform, Enterpraise II) placed in the vascular flow phantom were imaged with different the flip angle (FA) and echo time (TE). Each original image of the vascular flow phantom was reconstructed into a coronal multiplanar reconstruction (MPR) image and a profile curve along the long axis of the stent was calculated. The reduction of in-stent signal was assessed by relative in-stent signal (RIS) calculated by comparing intra-stent signal intensity to the reference tube in the original images. When the TE was 2.3 ms, the profile curve was the highest signal intensity. However, RIS had not changed by the extension of TE. When the FA was high, the RIS showed a high value, but when it exceeded 25 degrees, the signal attenuated in the distal part of the slab.The susceptibility artifacts and RF shielding generated around the stent can be reduced by the shortest TbE (2.3 ms) and 20 degrees of FA in the PCA imaging.
Related Concept Videos
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Magnetic Susceptibility and Permeability
When diamagnetic materials are placed under an external magnetic field, the moments opposite to the field are induced. Hence, the susceptibility for diamagnets has a minimal negative value of 10-5–10-6. Since...
Phase Diagrams
Diamagnetic Shielding of Nuclei: Local Diamagnetic Current
Wave Parameters
Reducing Line Loss
With a step-up transformer at the source, the voltage is increased, thereby reducing the current in the transmission lines since power loss in...