Reproducibility of inhomogeneous magnetization transfer (ihMT): A test-retest, multi-site study
Lei Zhang1, Tao Chen1, Hongzhe Tian1
1Department of Radiology, Baoji Center Hospital, Baoji, Shaanxi, People's Republic of China.
Background:
Inhomogeneous magnetization transfer (ihMT) has been reported to feature superior sensitivity and specificity for myelin imaging. However, the reproducibility on ihMT has yet been rarely investigated up to date. The purpose of the present study is to assess the multi-center reproducibility and test-retest variability of ihMT in central nervous system.
Methods:
5 volunteers were recruited and scanned twice on three 3.0 T magnetic resonance imaging (MRI) scanners using ihMT with identical scan parameters. The maps of quantitative ihMT (qihMT) and ihMT ratio (ihMTR) for each scan were calculated. Voxel based analysis then was performed to generate qihMT and ihMTR values for major white matter fibers. The intra- and inter-scanner reliability and reproducibility was assessed with intraclass correlation coefficients (ICCs). Bland-Altman method was used to show the level of agreement between two measurement types. Paired t-test and one-way ANOVA test were also used to compare the difference between inter- and intra-scanner, respectively.
Results:
In the 10 major white matter tracts areas, the ICCs indicated high intra- and inter-scanner measurement reliability and reproducibility. The Bland-Altman plots together with 95% confidence interval (CI) across all ROIs in the five volunteers also demonstrated good repeatability. No significant inter- and intra-scanner differences were found in Paired t-test and one-way ANOVA tests.
Conclusion:
Good inter- and intra-scanner reliability and reproducibility of ihMT measurements were observed in this study. These findings support the use of ihMT measurements as biomarkers in multicenter and/or longitudinal studies.
More Related Videos
09:12A Quantitative Fluorescence Microscopy-based Single Liposome Assay for Detecting the Compositional Inhomogeneity Between Individual Liposomes
Published on: December 13, 2019
06:54Combining 3D Magnetic Force Actuator and Multi-Functional Fluorescence Imaging to Study Nucleus Mechanobiology
Published on: July 5, 2022
Related Concept Videos
Types of Genetic Transfer Between Organisms
Imaging Studies IV: Magnetic Resonance Imaging
Colors and Magnetism
When atoms or molecules absorb light at the proper frequency, their electrons are excited to higher-energy orbitals. For many main group atoms and molecules, the absorbed photons are in the ultraviolet range of the electromagnetic spectrum, which cannot be detected by the human eye. For coordination compounds, the energy difference between the d orbitals often allows photons in the visible range to be absorbed and emitted, which is seen as colors by the human...
Multi-input and Multi-variable systems
In the absence of...
Magnetism
An individual magnetic pole cannot be isolated. No matter how small, every piece of a magnet contains a north pole and a south...
Transfer RNA Synthesis
Each of these chemical modifications is carried by a specific enzyme, post-transcription. All of these enzymes have unique base and site-specificity. Methylation, the most common chemical modification, is carried by at least nine different enzymes, with...