Dependency of prescribed CT dose on table height, patient size, and localizer acquisition for one clinical MDCT
James Winslow1, Yakun Zhang1, Lynne Koweek1
1Department of Radiology, Duke University Hospital, Durham, NC 27705, USA.
Purpose:
The purpose of this study was to quantify the effect that table height, patient size, and localizer acquisition order may have on AEC prescribed dose.
Method And Materials:
Three phantoms were used for this study: the Mercury Phantom, acrylic sheets, and an anthropomorphic phantom. A lateral (LAT) and a posterior-anterior (PA) localizer was acquired for each phantom at different table heights on a MDCT scanner (GE Discovery CT750 HD). AEC scan acquisitions were prescribed for each combination of phantom, localizer orientation, and table height ±4 cm with the center position; the displayed CTDIvol was recorded. Based on the institutional dose monitoring program, the relationship between change in CTDIvol and change in table height were studied for LAT and AP localizers for clinical exams.
Results:
For all phantom scans based on the PA localizer, the percent change in ranged between -18% and 42% for table heights 4 cm below and above proper centering; while for the LAT localizer, the percent change in CTDIvol from ideal were no greater than 12% different for ±4 cm differences in table height. Change in CTDIvol and change in table height displayed a strong linear relationship for AP localizer exams (P = 0.002), and weak correlation for LAT localizer exams (P = 0.12).
Conclusions:
Since uncertainty in vertical patient positioning is inherently greater than lateral positioning, the LAT localizer should be utilized to precisely and reproducibly deliver the intended amount of radiation prescribed by CT protocols.
More Related Videos
07:10Home-Based Prescribed Pulmonary Exercise in Patients with Stable Chronic Obstructive Pulmonary Disease
Published on: August 24, 2019
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Dose Size and Dosing Frequency: Determination Methods
Drug Dosing: Geriatric Patients
Drug Dosing: Obese Patients
Nonlinear Pharmacokinetics: Dependence of Elimination Half-Life and Dose Clearance
A study on guinea pigs examined the...
The Periodic Table
Flow Table Test
Concrete is placed within a truncated cone mold that is 8 inches high with an 8-inch base diameter and a 5-inch top diameter. The...