Comprehensive two-dimensional gas chromatography and time-of-flight mass spectrometry detection with a 50 ms
H Daniel Bahaghighat1, Chris E Freye2, Derrick V Gough2
1Department of Chemistry, Box 351700, University of Washington, Seattle, WA 98195, United States; Department of Chemistry and Life Science, United States Military Academy, West Point, New York, 10996, United States.
Abstract:
An ultrafast flow modulation period, PM of 50 ms, for comprehensive two-dimensional (2D) gas chromatography (GC × GC) with time-of-flight mass spectrometry (TOFMS) detection is demonstrated, producing narrow peak widths, 2W (4σ width-at-base on the 2D dimension), demonstrating the potential for ultrafast (2D) separations with high peak capacity. The modulator is a pulse flow valve that injects a narrow pulse of carrier gas at a user defined PM, at the union between the 1D and 2D columns. The raw data produced combines the properties of vacancy chromatography and frontal analysis. Deconvolution of the raw data using unconstrained multivariate curve resolution alternating least squares (MCR-ALS) analysis facilitates identification and quantification for overlapped analyte peaks. The peak profile loadings obtained from MCR-ALS are converted into traditional appearing GC × GC data through a process commonly used with frontal analysis. An 18-component test mixture at seven different injected mass levels was studied. The 2D peaks generated ranged from an 2W of 16 to 36 ms with an average of 26 ms. At an on-column injected mass of 14 ng per analyte, an average mass spectral match value, MV, of 822 was achieved using in-house collected spectra for comparison, with an average match value RSD of 7.1%. Calibration of overlapped test analytes was evaluated using the areas of the MCR-ALS loadings, with excellent quantification demonstrated. The advancement demonstrated in modulation performance for GC × GC represents a significant decrease in PM as most commercial modulators have a minimum PM of 1 s, while maintaining the benefits of a duty cycle of essentially 1.0, which promises to enable new chemical analyzer designs, compatible with the vacuum requirements of the TOFMS detector.
More Related Videos
08:56Detection of Regulated Ergot Alkaloids in Food Matrices by Liquid Chromatography-Trapped Ion Mobility Spectrometry-Time-of-Flight Mass Spectrometry
Published on: November 22, 2024
10:14Chromatographic Fingerprinting by Template Matching for Data Collected by Comprehensive Two-Dimensional Gas Chromatography
Published on: September 2, 2020
Related Concept Videos
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Applications of the Ideal Gas Law: Molar Mass, Density, and Volume