Determination of optimum exposure parameters for dentoalveolar structures of the jaws using the CB MercuRay system
Warangkana Weerawanich1,2, Mayumi Shimizu3, Yohei Takeshita4
1Department of Oral and Maxillofacial Radiology, Faculty of Dental Science, Kyushu University, 3-1-1 Maidashi, Higashi-ku, Fukuoka, 812-8582, Japan. yui@rad.dent.kyushu-u.ac.jp.
Objective:
To determine the optimum cone beam computed tomography exposure parameters for specific diagnostic tasks.
Methods:
A Teflon phantom attached to a half-mandible in a large container was scanned in dental (D), implant (I), and panoramic (P) modes. An identical phantom in a small container was scanned in D mode. Both were scanned at 60, 80, 100, and 120 kV. We evaluated the image quality of five anatomical structures [dentinoenamel junction (1), lamina dura and periodontal ligament space (2), trabecular pattern (3), cortex-spongy bone junction (4), and pulp chamber and root canal (5)] and analyzed the diagnostic image quality with cluster signal-to-noise analysis. We then evaluated correlations between the two image qualities and calculated the threshold of acceptable diagnostic image quality. Optimum exposure parameters were determined from images with acceptable diagnostic image quality.
Results:
For the small container, the optimum exposure parameters were D mode, 80 kV for (1), (3), and (4) and D mode, 100 kV for (5). For the large container, they were D mode, 120 kV for (1), (3), and (5) and D mode, 100 kV for (4). I mode, 120 kV reached the acceptable level for (4). No images reached the acceptable level for (2).
Conclusions:
No optimum exposure parameters were identified for the evaluation of the lamina dura and periodontal ligament space. D mode was sufficient for the other structures; however, the tube voltage required for each structure differed. Smaller patients required lower tube voltage. I mode, 120 kV may be used for larger lesions.
Related Concept Videos
Determination of Multiple Dosing Parameters: Loading and Maintenance Doses
Noncompartmental Analysis: Miscellaneous Pharmacokinetic Parameters
One key aspect of the noncompartmental approach is determining a drug's total clearance. This can be done by dividing the drug dose by the area under the concentration-time curve from zero to infinity. The area under the concentration-time curve represents the drug's...
Determination of Multiple Dosing Parameters: Steady-State, Minimum and Maximum Concentrations
Small-Signal Analysis of MOSFET Amplifiers
Small-Signal Analysis of BJT Amplifiers
Cluster Sampling Method
To choose a cluster sample, divide the population into clusters (groups) and then randomly select some of the clusters. All the members from these clusters are in the cluster sample. For example, if you randomly sample four departments from your...