Comparability of dental implant site ridge measurements using ultra-low-dose multidetector row computed tomography
Asma'a Abdurrahman Al-Ekrish1, Reema Al-Shawaf2, Wafa Alfaleh2
1Division of Oral and Maxillofacial Radiology, Department of Oral Medicine and Diagnostic Sciences, College of Dentistry, King Saud University, PO Box 60169, Riyadh, 11545, Saudi Arabia. aalekrish@ksu.edu.sa.
Objective:
To assess the linear measurements of edentulous ridges recorded from multidetector row computed tomography (MDCT) images obtained by a previously untested ultra-low dose in combination with filtered back-projection (FBP), adaptive statistical iterative reconstruction (ASIR), and model-based iterative reconstruction (MBIR).
Methods:
Three cadavers were imaged using a reference protocol with a standard dose and FBP (volume CT dose index (CTDIvol): 29.4 mGy) and two ultra-low-dose protocols, LD1 and LD2 (CTDIvol: 0.53 and 0.29 mGy). All test examinations were reconstructed with FBP, ASIR 50, ASIR 100, and MBIR. Linear measurements from the images of the edentulous ridges recorded from the test protocols were compared with those from the reference using a one-sample t test, Bland-Altman plots, and linear regression. Statistical significance was set at a p value of 0.05.
Results:
The one-sample t test demonstrated a statistically significant difference between the measurements from the reference protocol and all test protocols. The difference was not clinically significant for the following three test protocols: LD1/FBP, LD1/ASIR 50, and LD2/FBP. Bland-Altman plots with linear regression showed no systematic variation between the measurements obtained with the reference protocol and these three test protocols.
Conclusions:
The lowest-dose protocol to demonstrate comparable measurements with a standard MDCT dose was CTDIvol 0.29 mGy with FBP. These results must be considered with caution for areas of the jaws with thin cortication. The results in areas of thin cortication should be verified by studies with larger sample sizes at such areas and comparison with true gold standard measurements.
More Related Videos
Related Concept Videos
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Reconstruction of Signal using Interpolation
Imaging Studies III: Computed Tomography
Statistical Significance
Passive Filters
Low-Pass Filters
Low-pass filters are designed to transmit signals with frequencies lower than the cutoff frequency, ωc, and attenuate those above it. The cutoff...