Pre- and post-contrast versus post-contrast cone-beam breast CT: can we reduce radiation exposure while maintaining
Johannes Uhlig1, Uwe Fischer2, Lorenz Biggemann3
1Institute for Diagnostic and Interventional Radiology, University Medical Center Göttingen, Robert-Koch-Street 40, 37075, Göttingen, Germany. johannes.uhlig@med.uni-goettingen.de.
Objectives:
To evaluate whether post-contrast cone-beam breast CT (CBBCT) alone is comparable to the current standard of combined pre- and post-contrast CBBCT regarding diagnostic accuracy and superior regarding radiation exposure.
Material And Methods:
This study included 49 women (61 breasts) with median age 57.9 years and BI-RADS 4/5 lesions diagnosed on mammography/ultrasound in density type c/d breasts. Two radiologists rated post-contrast CBBCT and pre- and post-contrast CBBCT with subtraction images on the BI-RADS scale separately for calculation of inter- and intra-observer agreement and in consensus for diagnostic accuracy assessment. Sensitivity, specificity, and area under the curve (AUC) were compared via McNemar test and DeLong method, respectively. Subtraction imaging misregistration were measured from 1 (no artifacts) to 4 (artifacts with width > 4 mm).
Results:
A total of 100 lesion (51 malignant; 6 high risk; 43 benign) were included. AUC, sensitivity, and specificity showed no significant differences comparing post-contrast CBBCT alone versus pre- and post-contrast CBBCT (AUC 0.84 vs. 0.83, p = 0.643; sensitivity 0.89 vs. 0.85, p = 0.158; specificity 0.73 vs. 0.76, p = 0.655). Inter- and intra-observer agreement was excellent (intra-class correlation coefficient ICC = 0.76, ICC = 0.83, respectively). Radiation dose was significantly lower for post-contrast CBBCT alone versus pre- and post-contrast CBBCT (median average glandular radiation dose 5.9 mGy vs. 11.7 mGy, p < 0.001). High-degree misregistrations were evident in the majority of subtraction images (level 1/2/3/4 16.9%/27.1%/16.9%/39%), in particular for bilateral exams (3.2%/29.2%/8.3%/58.3%).
Conclusion:
Diagnostic accuracy of post-contrast CBBCT alone is comparable to pre- and post-contrast CBBCT in type c/d breasts, while yielding a significant twofold radiation dose reduction.
Key Points:
• The diagnostic accuracy of post-contrast CBBCT alone is comparable to dual acquisition of pre- and post-contrast CBBCT. • Acquisition of the post-contrast CBBCT scan alone reduces radiation exposure compared to pre- and post-contrast CBBCT, thus countering one of the main limitations of CBBCT. • High-degree misregistration artifacts limit the interpretation of subtraction images from pre- and post-contrast CBBCT studies.
Related Concept Videos
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Cardiomyopathy VII: Pre and Post Operative Nursing Management
Post-traumatic Stress Disorder
Symptoms and Behavioral Manifestations
A spectrum of distressing symptoms characterizes PTSD. Recurrent flashbacks, where individuals involuntarily relive traumatic events,...
Post-translational Translocation of Proteins to the RER
Targeting proteins to the ER
Hsp40 and Hsp70 chaperone molecules bind the translated proteins in the cytosol to prevent their folding. The chaperone binding helps to keep the signal...
Biological Effects of Radiation
Handwashing III: During the Procedure and Post-Procedure Steps