Dosimetric Impact of MRI Distortions: A Study on Head and Neck Cancers
Mary Adjeiwaah1, Mikael Bylund1, Josef A Lundman1
1Department of Radiation Sciences, Umeå University, Umeå, Sweden.
Purpose:
To evaluate the effect of magnetic resonance (MR) imaging (MRI) geometric distortions on head and neck radiation therapy treatment planning (RTP) for an MRI-only RTP. We also assessed the potential benefits of patient-specific shimming to reduce the magnitude of MR distortions for a 3-T scanner.
Methods And Materials:
Using an in-house Matlab algorithm, shimming within entire imaging volumes and user-defined regions of interest were simulated. We deformed 21 patient computed tomography (CT) images with MR distortion fields (gradient nonlinearity and patient-induced susceptibility effects) to create distorted CT (dCT) images using bandwidths of 122 and 488 Hz/mm at 3 T. Field parameters from volumetric modulated arc therapy plans initially optimized on dCT data sets were transferred to CT data to compute a new plan. Both plans were compared to determine the impact of distortions on dose distributions.
Results:
Shimming across entire patient volumes decreased the percentage of voxels with distortions of more than 2 mm from 15.4% to 2.0%. Using the user-defined region of interest (ROI) shimming strategy, (here the Planning target volume (PTV) was the chosen ROI volume) led to increased geometric for volumes outside the PTV, as such voxels within the spinal cord with geometric shifts above 2 mm increased from 11.5% to 32.3%. The worst phantom-measured residual system distortions after 3-dimensional gradient nonlinearity correction within a radial distance of 200 mm from the isocenter was 2.17 mm. For all patients, voxels with distortion shifts of more than 2 mm resulting from patient-induced susceptibility effects were 15.4% and 0.0% using bandwidths of 122 Hz/mm and 488 Hz/mm at 3 T. Dose differences between dCT and CT treatment plans in D50 at the planning target volume were 0.4% ± 0.6% and 0.3% ± 0.5% at 122 and 488 Hz/mm, respectively.
Conclusions:
The overall effect of MRI geometric distortions on data used for RTP was minimal. Shimming over entire imaging volumes decreased distortions, but user-defined subvolume shimming introduced significant errors in nearby organs and should probably be avoided.
More Related Videos
07:30A Test Bed to Examine Helmet Fit and Retention and Biomechanical Measures of Head and Neck Injury in Simulated Impact
Published on: September 21, 2017
05:45In Vitro Establishment of a Genetically Engineered Murine Head and Neck Cancer Cell Line using an Adeno-Associated Virus-Cas9 System
Published on: January 9, 2020
Related Concept Videos
Arteries of the Head and Neck
The internal carotid arteries supply blood to the anterior portion of the cerebrum. They enter the...
Veins of Head and Neck
On the other hand, the vertebral veins, unlike their arterial counterparts, are not primarily responsible for brain drainage. Instead, they drain the cervical vertebrae, spinal cord, and some small...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Muscles of the Anterior Neck
Impact of Groups on Groups
¹H NMR: Interpreting Distorted and Overlapping Signals
As Δν decreases and the signals move closer, the doublets appear increasingly distorted. The intensities of the inner lines increase at the cost of those of the outer lines as the signals are...