Fluorescence detection methods for microfluidic droplet platforms
Published on: December 10, 2011
Advancement of analytical modes in a multichannel, microfluidic droplet-based sample chopper employing phase-locked
Jean T Negou1, Juan Hu1, Xiangpeng Li1
1Department of Chemistry and Biochemistry, Auburn University, Auburn, AL 36849, USA.
Abstract:
In this work, we expand upon our recently developed droplet-based sample chopping concepts by introducing a multiplexed fluidic micro-chopper device (μChopper). Six aqueous input channels were integrated with a single oil input, and each of these seven channels was controlled by a pneumatic valve for automated sampling through software control. This improved design, while maintaining high precision in valve-based droplet generation at bandwidths of 0.03 to 0.05 Hz, enabled a variety of analytical modes to be employed on-chip compared to previous devices limited to sample/reference alternations. The device was analytically validated for real-time, continuous calibration with a single sample and five standards; multiplexed analysis during calibration using a mixed mode; and standard addition through spiking of six sample droplets with varying amounts of standard. Finally, the standard addition mode was applied to protein quantification in human serum samples using on-chip, homogeneous fluorescence immunoassays. Ultimately, with only ~1.2 μL of total analyzed solution volume- representing 100-fold and 75-fold reductions in reagent and serum volumes, respectively-we were able to generate full, six-point standard addition curves in only 1.5 min, and results correlated well with those from standard plate-reader equipment. This work thus exploited microfluidic valves for both their automation and droplet phase-locking capabilities, resulting in a micro-analytical tool capable of complex analytical interrogation modes on sub-microliter sample volumes while also leveraging drastic noise rejection via lock-in detection. The multichannel μChopper device should prove particularly useful in analyzing precious biological samples or for dynamic analyses at small volume scales.
Related Concept Videos
Sample Preparation for Analysis: Advanced Techniques
Acid digestion with strong acids is commonly used to dissolve inorganic materials that are insoluble (do not dissolve) in water. This method can be useful for...
Self-Locking Screw
A square-threaded screw jack carrying a load is considered self-locking if the screw retains its position even after the moment applied to it is removed.
What is a Mode?
There can be more than one mode in a data set if multiple values have the same highest frequency. For instance, suppose that the Statistics exam scores of 20 students are: 50; 53; 59; 59; 63; 63; 72; 72; 72; 72; 72; 76; 78; 81; 83; 84; 84; 84; 90; 93. Here, the mode is 72, as it occurs most frequently, five times.
A data set with two modes is called bimodal. For example,...
Transmission-based Precautions I: Contact, Enteric, and Droplets
Contact Precautions:
Contact precautions are the measures taken to prevent the transmission of infectious agents, especially epidemiologically important microorganisms such as MRSA or influenza, primarily transmitted through direct or indirect contact with an...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Development of Analytical Methods