Ability to detect endodontic complications using three different cone beam computed tomography units with and without
C Koç1, K Kamburoğlu2, G Sönmez2
1Department of Endodontics, Faculty of Dentistry, Başkent University, Ankara, Turkey.
Aim:
To assess observer performance in detecting endodontic complications using three different cone beam computed tomography (CBCT) units with and without the application of artefact reduction modes.
Methodology:
The study involved 40 freshly extracted human mandibular teeth (n = 10 per group) and divided randomly into four endodontic complication groups. Group 1) Instrument fracture; Group 2) Strip perforation; Group 3) Canal underfilling; and Group 4) Canal overfilling. Images of each tooth were obtained using three different CBCT units offering artefact reduction algorithms: the ProMax 3D Max, the Pax Flex 3D and the Dentri S. Four observers evaluated the images for the presence/absence of the four simulated endodontic complications. Weighted kappa coefficients and intra-class correlation coefficients (ICCs) were calculated to reveal the intra- and inter-observer agreement for each imaging mode, respectively. Receiver operating characteristic (ROC) analysis was used to evaluate the observers' performance. DeLong tests were used to compare the results for each image mode and observer using a significance level of α = 0.05.
Results:
In each of the four simulated endodontic complication groups, no significant differences were observed with and without application of artefact reduction for any of the three CBCT units tested. Only two significant differences were detected, and both were between the ProMax 3D Max at low mode AR and ProMax 3D Max without AR: observer 2 in group 1 (P = 0.0001) and observer 4 in group 4 (P = 0.0256).
Conclusion:
For each of the three CBCT units tested, application of artefact reduction for detecting endodontic complications is not recommended as a routine tool.
More Related Videos
05:37Cone Beam Intraoperative Computed Tomography-based Image Guidance for Minimally Invasive Transforaminal Interbody Fusion
Published on: August 6, 2019
09:49Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization
Published on: December 2, 2013
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
What is a Mode?
There can be more than one mode in a data set if multiple values have the same highest frequency. For instance, suppose that the Statistics exam scores of 20 students are: 50; 53; 59; 59; 63; 63; 72; 72; 72; 72; 72; 76; 78; 81; 83; 84; 84; 84; 90; 93. Here, the mode is 72, as it occurs most frequently, five times.
A data set with two modes is called bimodal. For example,...
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
Oxidation-Reduction Reactions
Measurement: Standard Units