EFFECTS OF THE SCAN RANGE ON RADIATION DOSE IN THE COMPUTED TOMOGRAPHY COMPONENT OF ONCOLOGY POSITRON EMISSION
Yusuke Inoue1, Kazunori Nagahara2, Hiroko Kudo2
1Department of Diagnostic Radiology, Kitasato University School of Medicine, Kitasato, Minami-ku, Sagamihara, Kanagawa, Japan.
Abstract:
We performed phantom experiments to investigate radiation dose in the computed tomography component of oncology positron emission tomography/computed tomography in relation to the scan range. Computed tomography images of an anthropomorphic whole-body phantom were obtained from the head top to the feet, from the head top to the proximal thigh or from the skull base to the proximal thigh. Automatic exposure control using the posteroanterior and lateral scout images offered reasonable tube current modulation corresponding to the body thickness. However, when the posteroanterior scout alone was used, unexpectedly high current was applied in the head and upper chest. When effective dose was calculated on a region-by-region basis, it did not differ greatly irrespective of the scan range. In contrary, when effective dose was estimated simply by multiplying the scanner-derived dose-length product by a single conversion factor, estimates increased definitely with the scan range, indicating severe overestimation in whole-body imaging.
More Related Videos
10:21Continuous Blood Sampling in Small Animal Positron Emission Tomography/Computed Tomography Enables the Measurement of the Arterial Input Function
Published on: August 8, 2019
05:32Multimodal Bioluminescent and Positronic-emission Tomography/Computational Tomography Imaging of Multiple Myeloma Bone Marrow Xenografts in NOG Mice
Published on: January 7, 2019
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies III: Computed Tomography
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Biological Effects of Radiation
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...